ESA620 Electrical Safety Analyzer Fluke Biomedical 3
Mains on applied part
test voltage
110 % of mains at 230 V for IEC 60601
100 % of mains for AAMI at 115 V per AAMI
100 % of mains at 230 V per 62353
Differential leakage
Ranges
10 µA to 199 µA
200 µA to 1999 µA
2 mA to 20 mA
Accuracy
± 10 % of reading ± (2 counts or 20 µA, whichever is greater)
Insulation resistance
Ranges/accuracy
0.5 M
Ω to 20 MΩ
± (2 % of reading + 2
counts)
20 M
Ω to 100 MΩ
± (7.5 % of reading + 2
counts)
Source test voltage
500 V dc
250 V dc
Insulation resistance tests
Mains-PE, AP-PE, Mains- PE, Mains-NE (non-earthed accessible
conductive part) and AP- NE (non-earthed accessible conductive
part)
Mamimum load capacitance
1 µF
ECG performance waveforms
Accuracy
± 2 %
± 5 % for amplitude of 2 Hz square wave only, fixed at 1 mV
Lead II configuration
Waveforms
Rates
ECG complex (BPM)
30, 60, 120, 180, and 240
Ventricular fibrillation
Square wave (50 % duty cycle) (Hz)
0.125 and 2
Sine wave (Hz)
10, 40, 50, 60, and 100
Triangle wave (Hz)
2
Pulse (63 ms pulse width)
30 and 60
Power ratings
Mains voltage outlet
120 V ac
230 V ac
Mains voltage inlet power range 90 to 132 V ac rms
180 to 264 V ac rms
Maximum current
20 A
16 A
Hz
50 or 60
50 or 60
Physical case
Dimensions (L x W x H)
31 cm x 23 cm x 10 cm (12.2 in x 9 in x 2.9 in)
Weight
4.7 kg (10.25 lb)
Environmental specifications
Operating temperature
10 °C to 40 °C
Storage temperature
-20 °C to 60 °C
Operating humidity
10 % to 90 % non-condensing
Altitude
To 2,000 meters
General
Warranty
Two-year extended warranty*
*No-cost extended warranty available after first-year calibration at any Fluke Biomedical authorized service center.