ESA620 Electrical Safety Analyzer Fluke Biomedical 3

Mains on applied part
test voltage

110 % of mains at 230 V for IEC 60601
100 % of mains for AAMI at 115 V per AAMI
100 % of mains at 230 V per 62353

Differential leakage

Ranges

10 µA to 199 µA
200 µA to 1999 µA
2 mA to 20 mA

Accuracy

± 10 % of reading ± (2 counts or 20 µA, whichever is greater)

Insulation resistance

Ranges/accuracy

0.5 M

Ω to 20 MΩ

± (2 % of reading + 2
counts)

20 M

Ω to 100 MΩ

± (7.5 % of reading + 2
counts)

Source test voltage

500 V dc
250 V dc

Insulation resistance tests

Mains-PE, AP-PE, Mains- PE, Mains-NE (non-earthed accessible
conductive part) and AP- NE (non-earthed accessible conductive
part)

Mamimum load capacitance

1 µF

ECG performance waveforms

Accuracy

± 2 %
± 5 % for amplitude of 2 Hz square wave only, fixed at 1 mV
Lead II configuration

Waveforms

Rates

ECG complex (BPM)

30, 60, 120, 180, and 240

Ventricular fibrillation

Square wave (50 % duty cycle) (Hz)

0.125 and 2

Sine wave (Hz)

10, 40, 50, 60, and 100

Triangle wave (Hz)

2

Pulse (63 ms pulse width)

30 and 60

Power ratings

Mains voltage outlet

120 V ac

230 V ac

Mains voltage inlet power range 90 to 132 V ac rms

180 to 264 V ac rms

Maximum current

20 A

16 A

Hz

50 or 60

50 or 60

Physical case

Dimensions (L x W x H)

31 cm x 23 cm x 10 cm (12.2 in x 9 in x 2.9 in)

Weight

4.7 kg (10.25 lb)

Environmental specifications

Operating temperature

10 °C to 40 °C

Storage temperature

-20 °C to 60 °C

Operating humidity

10 % to 90 % non-condensing

Altitude

To 2,000 meters

General

Warranty

Two-year extended warranty*

*No-cost extended warranty available after first-year calibration at any Fluke Biomedical authorized service center.