Atec Hipotronics-HD100 Series User Manual
Page 2
SPECIFICATIONS
HD100
Panel Detail
METERS
OVERLOAD SENSITIVITY
OUTPUT CURRENT
METER RANGE
TERMINATIONS
INPUT VOLTAGE
MODEL
HD103
HD106
HD115
HD125
HD140
DIMENSIONS
* DC output voltage is negative with respect to ground.
4-1/2" meter, ± 2% full scale accuracy
Adjustable from 0.5 mA to 5.5 mA
0-5 mA ac, 0-50/500/5000 µA dc
High voltage lead – 5 ft. long shielded high voltage test lead of RG58/U (RG8/U for
HD140) with alligator clip. Ground lead – 5 ft. long insulated test lead with alligator clip.
Model number suffix -A
120 V / 60 Hz
Model number suffix -B
220 V / 50 Hz
21"W x 19"D x 11"H (54 x 48 x 28 cm)
ACCESSORIES
• Safety Test Cage – The interlocked test cage door shuts off high voltage if the door
opens. Catalog No. TC-25.
• Retractable Test Probe – The spring loaded, pistol-like probe allows exposure of
high voltage only when the voltage is to be applied to the object under test (not
rated for use with the HD115, HD125 or HD140). Catalog No. HTP-R.
• Fixed Test Probe – Allows easy application of the high voltage to the object under
test. Rated for use with all models except the HD140. Catalog No. HTP-F.
• Spare Parts Kits – Catalog Nos. SPK1-HD103, SPK1-HD106, SPK1-HD115,
SPK1-HD125 & SPK1-HD140
For further information, contact:
Hipotronics, Inc.
A Subsidiary of Hubbell, Incorporated
Route 22, P. O. Box 414
Brewster, NY 10509, U.S.A.
1-800-727-4476
Tel: 845-279-8091
Fax: 845-279-2467
SHIP WEIGHT
59 lb (27 kg)
64 lb (29 kg)
85 lb (39 kg)
85 lb (39 kg)
92 lb (47 kg)
NET WEIGHT
46 lb (21 kg)
51 lb (23 kg)
75 lb (34 kg)
75 lb (34 kg)
82 lb (37 kg)
VOLTMETER RANGE
0-0.6/1.2/3.0 kV
0-1.2/3.0/6.0 kV
0-3.75/7.5/15.0 kV
0-5.0/10.0/25.0 kV
0-8.0/16.0/40.0 kV
OUTPUT VOLTAGE*
0-2.5 kV ac
0-3.0 kV dc
0-5.0 kV ac
0-6.0 kV dc
0-12.5 kV ac
0-15.0 kV dc
0-10.0 kV ac
0-25.0 kV dc
0-15.5 kV ac
0-40.0 kV dc
NOVEMBER 1995
HD100-DS-1