Phenix, Safety & design features description, Environmental conditions – Atec Phenix-6CP Series User Manual
Page 2
TECHNOLOGIES
PHENIX
R
www.phenixtech.com
SAFETy & DESIGN FEATURES
DESCRIPTION
This portable rugged line of AC high voltage test equipment is
ideally suited for use in the field or in the laboratory. The 6CP line
offers a complete test system in a mobile package. All units in the
6CP line use a combination control and regulator enclosure and
a separate high voltage section. The system is equipped with
advanced digital metering, and control and protection devices
which are easy to use and provide the operator with all necessary
information to safely operate the system. The Metering Guard Mode
allows accurate measurements of the high voltage test specimen
leakage current by bypassing stray capacitive currents. The high
voltage section consists of a specially designed oil insulated high
voltage transformer characterized by use of an isolating cylindrical
housing mounted on a rugged steel base. The high voltage output
is equipped with a corona free aluminum shielding electrode. The
high quality design and construction is key to the superior partial
discharge performance of the 6CP line. The system arrives ready
to use with all power and interconnect cables included; the simple
setup allows this 6CP to be ready for testing in minutes. The rugged
construction and dependability, backed by long-term warranty
makes the 6CP line an ideal choice for your high voltage testing
requirements.
•
2 or 3 piece designs
•
Circuit breaker protection
•
External interlock provision
•
High voltage On/Off pushbuttons
with indicator
•
Zero Start interlock
•
Manual control of output voltage
•
Adjustable
overload circuit
•
Slow and fast-acting transient
protection on all meters and relays
•
Digital timer with audible alarm
•
Digital multi range metering
•
Emergency Off pushbutton with
key lockout
•
Guard mode for accurate current
measurements
•
Output overload indicator with
reset switch
•
All interconnect control and power
cables are included
•
Cable storage provision
•
Two copies of operation/
maintenance manual
ENVIRONMENTAL CONDITIONS
•
10-40°C, indoor/outdoor in fair weather
•
Humidity <95% non-condensing
•
Altitude <3000 ft (1000 meters)
50% Inductive Reactive 50% Inductive Reactive
INPUT
OUTPUT
FEATURES
PD
DUT
y
C
yCLE
S
CABLES INCLUDED
MODEL
6CP30/15-3 6CP50/10-3 6CP100/50-7.5 6CP120/60-7.5 6CP100/50-10 6CP120/60-10 6CP200/100-10
110-120 V, 16 A 110-120 V, 16 A 110-120 V, 30 A 110-120 V, 30 A
220-240 V, 8 A 220-240 V, 8 A 220-240 V, 15 A 220-240 V, 15 A
Frequency
50/60 Hz
50/60 Hz
50/60 Hz
50/60 Hz
50/60 Hz
50/60 Hz
50/60 Hz
≈ 0-15 kV, 200 mA ≈ 0-10 kV, 300 mA ≈ 0-50 kV, 150 mA ≈ 0-60 kV, 125 mA ≈ 0-50 kV, 200 mA ≈ 0-60 kV, 167 mA ≈ 0-100 kV, 100 mA
≈ 0-30 kV, 100 mA ≈ 0-50 kV, 60 mA ≈ 0-100 kV, 75 mA ≈ 0-120 kV, 62.5 mA ≈ 0-100 kV, 100 mA ≈ 0-120 kV, 83 mA ≈ 0-200 kV, 50 mA
5 min ON/15 min OFF
3 kVA
7.5 kVA
10 kVA
15 min ON/1 hr OFF
2 kVA
5 kVA
7.5 kVA
1 hr ON/1 hr OFF
1.5 kVA
3 kVA
5 kVA
66% Inductive + Selectable
Capacitive Preload
Metering Accuracy
0.8% Reading +.2% Full Scale
Voltmeter Range(s)
0-15/30 kV
0-10/50 kV
0-50/100 kV
0-60/120 kV
0-50/100 kV
0-60/120 kV
0-100/200 kV
0-200 µA/2 mA/ 0-200 µA/2 mA/
20 mA/200 mA 20 mA/300 mA
Output
HV Electrode
PD level
<10 pC
Optional up to 100kV
<3 pC
Power Cable
10’(3048 mm)
30’(9144 mm)
Power Interconnect
10’(3048 mm)
20’(6096 mm)
Meter Interconnect
10’(3048 mm)
20’(6096 mm)
Ground
10’(3048 mm)
(2) 20’(6096 mm) each
(3) 30’(9144 mm) ea
Return
10’(3048 mm)
20’(6096 mm)
10’(3048 mm)
Divider Metering
n/a
n/a
30’(9144 mm)
Paralleling
n/a
n/a
10’(3048 mm)
Voltage/Current
Voltage/Current
Compensation
0-200 µA/ 2 mA/ 20 mA/ 200 mA
Currentmeter Range(s)
220-240 V, 30 A 220-240 V, 30 A
220-240 V, 25 A