Atec OMICRON-CP-TC12 User Manual
Cp tc12 - 12 kv oil test cell, Accessory to cpc 100
mtronics technology
Technical Data
Cell Type
Three-terminal, guarded
Cell Gap Spacing
11 mm / 0.43 in
Meas. Electrode (Ø x H)
149 x 104 mm / 5.9 x 4.1 in
Sample Volume
1.2 - 2 liters / 41 - 68 fl.oz.
Cell Capacitance
65 pF, nominal (in air)
Maximum Test Voltage
12 kV rms
Voltage Operating Range 2.5 - 12 kV for a stress of
200 - 1100 V/mm / 7.9 - 43.3 V/in
HV Connectors
6 mm sockets
Meas. Connectors
4 mm sockets
Inner Dimensions (Ø x H) 172 x 180.8 mm / 6.8 x 7.1 in
Outer Dimensions
(W x H x D)
220 x 235.5 x 220 mm
8.7 x 9.3 x 8.7 in
Weight (approx.)
9.2 kg / 20 lbs
Order Number
VEHZ0601
CP TC12 - 12 kV Oil Test Cell
Accessory to CPC 100
The CP TC12 is used in conjunction with the CPC 100
and CP TD1 to measure tangent delta / power factor
and permittivity of insulation liquids, e.g. transformer
oil. The three-electrode design with guard allows precise
measurement, especially of small losses.
The circular electrodes are constructed from rigid
stainless steel and require a sample of 1.2 to 2 liters /
41 to 68 fl.oz.
Electrical connection to the Test Cell is made using the
standard cables provided with the CP TD1.