Atec Omicron-CP-CB2 User Manual
Cp cb2 - current booster, Accessory to cpc 100
CP CB2 - Current Booster
Accessory to CPC 100
Tests applications up to 2000 A
The output current of the CPC 100 gets increased to up to
2000 A. The CP CB2 can be connected with the short high
current cable set close to the busbar and with the long
connection cable to the CPC 100.
1 With mains voltage 230 V at 23 °C ± 5 °C (73 F ± 10 F) ambient temperature.
2 Signals below 50 Hz or above 60 Hz with reduced values possible.
3 Outputs in series.
4 Outputs in parallel.
Current Outputs
1000 A AC
3
2000 A AC
4
Amplitude
0 ... 500 A
0 ... 1000 A
0 ... 2000 A
tmax
1
30 min
25 s
25 s
Vmax
2
5.00 V
4.90 V
2.45 V
Pmax
2
2500 VA
4900 VA
4900 VA
f
15 - 400 Hz
15 - 400 Hz
15 - 400 Hz
Typical Accuracy
2
1000 A AC
2000 A AC
Amplitude
Reading
0.13 %
0.13 %
Full Scale
0.13 %
0.13 %
Phase
Full Scale
0.25 °
0.25 °
Guaranteed Accuracy
2
1000 A AC
2000 A AC
Amplitude
Reading
0.25 %
0.25 %
Full Scale
0.25 %
0.25 %
Phase
Full Scale
0.50 °
0.50 °
Technical Data
Output Current
up to 2000 A
Dimension
186 x 166 x 220 mm
7.3 x 6.5 x 8.7 in
Weight
16.0 kg / 35.3 lbs
Order Number
VEHZ 0630
Scope of Delivery
CP CB2 Current Booster
+ Grounding cable
+ Transport case
+ High current cable set
(Order number: VEHK 0610)
4 x 95 mm
•
2
with plugs and
clamps, 1.5 m / 5 ft
1 x 95 mm2 to serialize
•
outputs, 0.6 m / 2 ft
+ Connection cable to CPC 100
(Order number: VEHK 0611)
3 x 2.5 mm
•
2