beautypg.com

Atec Omicron-CP-CB2 User Manual

Cp cb2 - current booster, Accessory to cpc 100

background image

CP CB2 - Current Booster

Accessory to CPC 100

Tests applications up to 2000 A

The output current of the CPC 100 gets increased to up to
2000 A. The CP CB2 can be connected with the short high
current cable set close to the busbar and with the long
connection cable to the CPC 100.

1 With mains voltage 230 V at 23 °C ± 5 °C (73 F ± 10 F) ambient temperature.
2 Signals below 50 Hz or above 60 Hz with reduced values possible.
3 Outputs in series.
4 Outputs in parallel.

Current Outputs

1000 A AC

3

2000 A AC

4

Amplitude

0 ... 500 A

0 ... 1000 A

0 ... 2000 A

tmax

1

30 min

25 s

25 s

Vmax

2

5.00 V

4.90 V

2.45 V

Pmax

2

2500 VA

4900 VA

4900 VA

f

15 - 400 Hz

15 - 400 Hz

15 - 400 Hz

Typical Accuracy

2

1000 A AC

2000 A AC

Amplitude

Reading

0.13 %

0.13 %

Full Scale

0.13 %

0.13 %

Phase

Full Scale

0.25 °

0.25 °

Guaranteed Accuracy

2

1000 A AC

2000 A AC

Amplitude

Reading

0.25 %

0.25 %

Full Scale

0.25 %

0.25 %

Phase

Full Scale

0.50 °

0.50 °

Technical Data

Output Current

up to 2000 A

Dimension

186 x 166 x 220 mm
7.3 x 6.5 x 8.7 in

Weight

16.0 kg / 35.3 lbs

Order Number

VEHZ 0630

Scope of Delivery

CP CB2 Current Booster
+ Grounding cable
+ Transport case
+ High current cable set

(Order number: VEHK 0610)

4 x 95 mm

2

with plugs and

clamps, 1.5 m / 5 ft
1 x 95 mm2 to serialize

outputs, 0.6 m / 2 ft

+ Connection cable to CPC 100

(Order number: VEHK 0611)

3 x 2.5 mm

2

, 20 m / 66 ft