Atec Fluke 62MAX User Manual
Heat, Takes the
62 Max
62 Max+
Temperature Range
-30 °C to 500 °C (-22 °F to 932 °F)
-30 °C to 650 °C (-22 °F to 1202 °F)
Accuracy
± 1.5 °C or ± 1.5 % of reading
± 2.0 at -10 °C to 0 °C
± 3.0 at -30 °C to -10 °C
± 1. °C or ± 1.0 % of reading
± 2.0 at -10 °C to 0 °C
± 3.0 at -30 °C to -10 °C
Response Time (95 %)
<500 ms (95 % of reading)
Spectral Response: 8 to 14 microns
Emissivity: 0.10 to 1.00
<300 ms (95 % of reading)
Spectral Response: 8 to 14 microns
Emissivity: 0.10 to 1.00
Optical Resolution
10:1 (calculated at 90 % energy)
12:1 (calculated at 90 % energy)
Display Resolution
0.1 °C (0.2 °F)
0.1 °C (0.2 °F)
Repeatability of Readings
± 0.8 % of reading or <± 1.0 °C (2 °F),
whichever is greater
± 0.5 % of reading or <± 0.5 °C (1 °F),
whichever is greater
Power
AA battery
AA battery
TAKES THE
HEAT
.
TAKES THE
HEAT
.
AND THE DUST, WATER
AND 3-METER DROP.
AND THE DUST, WATER
AND 3-METER DROP.
Introducing the new 62 Max
IR Thermometers from Fluke.
General Specifications
Fluke Corporation
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40
2675 200 or Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2012 Fluke Corporation.
Specifications subject to change without notice.
Printed in U.S.A. 3/2012 4160018A_EN
Modification of this document is not permitted
without written permission from Fluke Corporation.
Learn more: fluke.com/62MAX