beautypg.com

Detailed specifications – Atec Fluke-Ti 200-300-400 User Manual

Page 2

background image

Detailed specifications

Ti400

Ti300

Ti200

Temperature
Temperature measurement range

(not calibrated below -10 °C)

-20 °C to +1200 °C

(-4 °F to +2192 °F)

-20 °C to +650 °C

(-4 °F to +1202 °F)

Temperature measurement

accuracy

± 2 °C or 2 % (at 25 °C nominal, whichever is greater)

On-screen emissivity correction

Yes (by number and table)

On-screen reflected background

temperature compensation

Yes

On-screen transmission correction

Yes

Imaging performance
Image capture frequency

9 Hz refresh rate or 60 Hz refresh rate depending upon model variation

Detector type

Focal Plane Array, uncooled

microbolometer, 320 x 240 pixels

Focal Plane Array, uncooled

microbolometer, 240 X 180 pixels

Focal Plane Array, uncooled

microbolometer, 200 X 150 pixels

Thermal sensitivity (NETD)

≤ 0.05 °C at 30 °C target temp (50 mK)

≤ 0.075 °C at 30 °C target temp (75 mK)

Total pixels

76,800

43,200

30,000

Infrared spectral band

7.5 μm to 14 μm (long wave)

Visual (visible light) camera

Industrial performance 5.0 megapixel

Standard infrared lens type

Field of view

24 ° x 17 °

Spatial resolution (IFOV)

1.31 mRad

1.75 mRad

2.09 mRad

Minimum focus distance

15 cm (approx. 6 in)

Optional telephoto infrared lens type, available soon

Field of view

12 ° x 9 °

Spatial resolution (IFOV)

0.65 mRad

0.87 mRad

1.05 mRad

Minimum focus distance

45 cm (approx. 18 in)

Optional wide-angle infrared lens type, available soon

Field of view

46 ° x 34 °

Spatial resolution (IFOV)

2.62 mRad

3.49 mRad

4.19 mRad

Minimum focus distance

15 cm (approx. 6 in)

Focus mechanism
LaserSharp

Auto Focus System

Yes

Advanced Manual Focus

Yes

Image presentation
Palettes

Standard

Ironbow, Blue-Red, High Contrast, Amber, Amber Inverted, Hot Metal, Grayscale, Grayscale Inverted

Ultra Contrast™

Ironbow Ultra, Blue-Red Ultra, High Contrast Ultra, Amber Ultra, Amber Inverted Ultra,

Hot Metal Ultra, Grayscale Ultra, Grayscale Inverted Ultra

Level and span

Smooth auto-scaling and manual scaling of level and span

Fast auto toggle between manual

and auto modes

Yes

Fast auto-rescale in manual mode

Yes

Minimum span (in manual mode)

2.0 °C (3.6 °F)

Minimum span (in auto mode)

3.0 °C (5.4 °F)

IR-Fusion

®

information

Picture-In-Picture (PIP)

Yes

Full screen infrared

Yes

AutoBlend™ mode

Yes

Color alarms (temperature alarms)

High-temperature , low-temperature, and isotherm (user-selectable)

Image capture and data storage
Image capture, review,

save mechanism

One-handed image capture, review, and save capability

Storage medium

Micro SD Memory Card, on-board flash memory, save-to-USB capability, direct download via USB-to-PC connection

File formats

Non-radiometric (.bmp) or (.jpeg) or fully-radiometric (.is2) Video*: non-radiometric (MPEG - encoded .AVI) and fully-radiometric (.IS3)

No analysis software required for non-radiometric (.bmp, .jpg and .avi*) files

Export file formats w/SmartView

®

software

BMP, DIB, GIF, JPE, JFIF, JPEG, JPG, PNG, TIF, and TIFF

Memory review

Thumbnail view navigation and review selection

Other time-saving and productivity features
Voice annotation

60 seconds maximum recording time per image; reviewable playback on imager

IR-PhotoNotes™

Yes

Wi-Fi connectivity

Yes, to PC, iPhone

®

, iPad

®

and WiFi to LAN*

Text annotation*

Yes

Video recording*

Standard and Radiometric

Streaming Video

Via USB to PC and HDMI to HDMI compatible screen

CNX™ Wireless System*

Yes*

Cardinal Compass*

Yes*

GPS coordinates and recording

Yes

Auto capture

(temperature and interval)

*

Yes*

Remote control and operation (for

special and advanced applications)

Yes

No

No

* Coming soon via firmware update. Users notified via SmartView software when available.

2 Fluke Corporation Ti400, Ti300, and Ti200 Industrial Thermal Imagers