TECHNICAL SPECIFICATIONS

PD Measuring

Input Impedance

50

PD System Bandwidth

30 kHz … 1.5 MHz (-6dB)
(with internal analog filter)

10 kHz … 1.5 MHz (-6dB)
(without internal analog filter)

Filter Center Frequency Freely selectable (32 kHz … 1.498

MHz)

Filter Bandwidth

4 kHz / 4.5 kHz / 9 kHz

10 … 100 kHz, in 10 kHz steps

100 … 500 kHz, in 50 kHz steps

600 kHz … 1 MHz, in 100 kHz steps

Sensitivity

< 0.1 pC (directly at AKV 9310/T
quadripole input)

Input Atten. Range

0 dB / 20 dB / 40 dB

Linearity Error

<

5 % (1 … 100 % FSR)

Pulse Phase Resolution 0.35°

Voltage Measurement
Input Voltage Range

0.14 … 140 VAC (RMS)

-200 … -0.2 / +0.2 … +200 VDC

Frequency Range

15 … 400 Hz

Input Impedance

1.7 M

/ 11 pF

Linearity Error

<

1 % (0.1 … 100 % FSR)

Synchronization

Voltage or Mains input

Synchronization Accur.

< 5°

Environmental
Operating temperature

0°C … +45°C

Storage temperature

-20°C ... +60°C

Relative Humidity

5% … 80%, non-condensing


RIV Measurement System
Frequency Range

850 … 1150 kHz

IF Filter Bandwidth

4.5 kHz and 9 kHz

Image Rejection

> 55 dB

Stop Band Attenuation

> 70 dB @

7.5 kHz

> 60 dB @

10 kHz

Linearity Error

2 dB @ 10 Hz … 1 kHz PRF, 1/10

… 10x calibration level

Quasi-Peak Detector
Response

According to NEMA 107, ANSI
C63.2-1996


Power Supply & Mechanical
Voltage

90 … 264 VAC

Frequency

50 / 60Hz

Dimensions (WxDxH)

483 mm x 306 mm x 89 mm

Weight

6.2 kg

EMC

Compliant to EMC directive
2004/108/EC (EN 61326-1)

Safety

Low voltage directive 2006/95/EC
(IEC 60010-1)

Vibration

MIL-STD-810G Table 514.6C-II,
Category common carrier

Applicable Standards
IEC-60060 Parts 1&2

ICEA T-24-380

IEC-60270

ASTM D1868-93

IEC-885-2 and 885-3

ANSI C57.113

IEEE Std. 4, 1995

ANSI C57.124-91

ANSI C63.2-1996

NEMA 107

ACCESSORIES FOR PD TESTING

CALIBRATORS

KAL 9520 High precision PD calibrator

High

precision

partial

discharge

calibrator touch screen controlled with
a very wide, continuously variable
output pC range

9216 hand held battery PD calibrator

Small battery powered PD calibrator.
Pulses outputs ranges are 10, 100, 1000
and 10’000 pC

KAL 9530 RIV calibrator

RIV calibrator (KAL 9530) including a
signal generator, RIV calibration set,
clamp and a switching.

MEASURING IMPEDANCES

AKV 9310 passive universal coupling impedance

Compact impedance fully passive
with enhanced sensitivity.

AKV 9330 Capacitors optimized coupling impedance

Active measuring impedance optimized
for PD testing of power capacitor as per
IEC 60270. It is rated for 300 amperes.

COUPLING CAPACITORS

9230 coupling capacitors

Consist of a high voltage unit mounted
on a mobile (with wheels) base frame
and includes a suitable top electrode
and a coupling impedance.

www.haefely.com

Europe, Asia, South &
Central America, Australia

Haefely Test AG
Birsstrasse 300
4052 Basel
Switzerland
 + 41 61 373 4111
 + 41 61 373 4912
[email protected]

China (Sales & Service Office)

Haefely Test AG – Beijing Office

8-1-602, Fortune Street

No. 67, Chaoyang Road, Chaoyang District

Beijing, 100025

P. R. China

 + 86 10 8578 8099

 + 86 10 8578 9908

[email protected]

North America

Hipotronics Inc.

1650 Route 22

PO Box 414

Brewster, NY 10509

USA

 + 1 845 279 3644

 + 1 845 279 2467

[email protected]