7400/7600
Measured Parameters:
Any two of 14 parameters measured and
displayed simultaneously, user selectable
Test Frequency:
7400
Range: 10 Hz to 500 kHz, continuous
Resolution: 0.1 Hz from 10 Hz to
10 kHz, 5 digits>10kHz
Accuracy: +/-(0.002% + 0.02Hz)
7600
Range: 10 Hz to 2 MHz, continuous
Resolution: 0.1 Hz from 10 Hz to
10 kHz, 5 digits>10kHz
Accuracy: +/-(0.25% + 0.02Hz)
Measurement Speed:
7400
Basic: 40 meas/sec
Enhanced: 8 meas/sec
Extended: 1 meas/sec
7600
Fast: 25 meas/sec
Medium: 8 meas/sec
Slow: 1 meas/sec
Ranging:
Automatic, Range Hold or user selectable
Trigger:
Internal (automatic), External (RS-232,
IEEE-488.2 or Handler interfaces) and Manual
AC Test Signal:
Voltage: 20 mV to 5.0 V (open circuit)
up to 500kHz in 5 mV steps
7600
20 mV to 1.0 V (open circuit)
500kHz-1MHz in 5 mV steps
20 mV to 0.5 V (open circuit)
>1MHz in 5 mV steps
Current: 250 µA to 100 mA (short circuit) in
50 µA steps
(Max. Compliance 3V < 500kHz.
Source Impedance:
25
Ω
, 400
Ω
, 6.4k
Ω
, or 100k
Ω
, range dependent
DC Bias Voltage:
Internal: 2.0 V
External: 0 to +/-200V
0 to +/-500V on
7400A/7600A
Display:
LCD Graphics with back light and adjustable
contrast
Result Formats:
Engineering or scientific notation
% Deviation from nominal of primary parameter
Deviation from nominal of primary parameter
Pass/Fail
Binning summary
No Display for maximum throughput
Sweep Result:
Primary parameter vs. frequency, voltage or
current
Graphical or Tabular Format
Up to 200 measurement points per sweep
Sequencing Result:
Displays up to 6 sequential test results, primary
and/or secondary
AutoAcc:
Automatic calculation and display of overall
instrument accuracy for selected settings, test
conditions and device under test
Standard Interfaces:
IEEE-488.2, RS-232, Handler, Printer Port,
3.5" Disk Drive
Charged Capacitor Protection:
√
8/C for Vmax
≤
250 V;
√
2/C for Vmax
≤
1000V
C = Capacitance in farads of the device under test
Additional Fuse Protection on
7400C
Measurement Delay:
Programmable from 0 - 1000 ms in 1 ms steps
Averaging:
Programmable from 1 - 1000
Median value mode
Data Storage:
40,000 measurements/disk
ASCII format
DOS compatible
Program Storage:
125 per disk
25 setups internal
Password protected
ASCII format
Calibration:
Recommended Calibration Interval 1 year
Complete NIST Traceable Calibration using
QuadTech 7000-09 Cal Kit
Built-in automatic calibration procedure
Usage & Cal Data:
Displays last calibration date, standard values
used in calibration and # of hours operation
Self-Test Routine:
Verifies critical instrument operation at power-
up or when selected from menu
Contact Check:
Time to detect, 2ms
Test Terminals:
Front panel, four terminal (BNC) guarded
Mechanical:
Bench mount with tilt bail
Dimensions: (w x h x d): 16 x 6 x 14in
(410 x 150 x 360mm)
Weight: 17 lbs (8kg) net, 23 lbs (10.5kg)
shipping
Environmental:
Meets MIL-T-28800E, Type 3, Class 5,
Style E & F
Operating: 0 to + 50˚ C
Humidity: <75% for 11 to 30˚ operating
Storage: - 40 to + 71˚ C
Power:
• 90 - 250Vac • 47/63Hz
• 100W max
For more detailed specifications, visit
www.quadtech.com •
For more information about special purchase, rent & lease options, call
1-800-253-1230
•
Fax 1-508-485-0295
•
Intl. 1-508-229-0806
5 Clock Tower Place, 210 East, Maynard, MA 01754
Pr
oduct Specif
ications
PN030087/A1
7400/7600 Precision LCR Meters
7600
LCR Meter
7600-CE
LCR Meter, CE Marked
7600A
LCR Meter, +/-500V External Bias
7600A-CE LCR Meter, +/-500V External Bias,
CE Marked
7400
LCR Meter
7400-CE
LCR Meter, CE Marked
7400A
LCR Meter, +/-500V External Bias
7400A-CE LCR Meter, +/-500V External Bias,
CE Marked
7400C
LCR Meter, Charged Capacitor
Protection
7400C-CE LCR Meter, Charged Capacitor
Protection, CE Marked
Includes:
Instruction Manual
Power Cord
Calibration Certificate Traceable to NIST
Optional Accessories:
7000-00
Rack Mount Kit
7000-01
BNC Cable Set, 1 meter
7000-02
BNC Cable Set, 2 meters
7000-03
Kelvin Clip Leads
7000-04
Alligator Clip Leads
7000-05
Chip Component Tweezers
7000-06
Axial/Radial Lead Component
Test Fixture
7000-07
Chip Component Test Fixture
7000-08
High Voltage Test Fixture
7000-09
Calibration Kit
Ordering Information
Cs, Cp
00000.01 fF to 9.999999 F
±0.5%
±0.25%
±0.05%
Ls, Lp
0000.001 nH to 99.99999 H
±0.5%
±0.25%
±0.05%
D
.0000001 to 99.99999
±0.005
±0.0025
±0.0005
Q
.0000001 to 999999.9
±0.005
±0.0025
±0.0005
Z , Rs, Rp,
000.0001 m
Ω
to 99.99999 M
Ω
±0.5%
±0.25%
±0.05%
ESR, Xs
Y , Gp, Bp
00000.01 µS to 9.999999 MS
±0.5%
±0.25%
±0.05%
Phase Angle
-180.0000 to +179.9999 degrees
±1.8˚
±0.9˚
±0.18˚
Basic Measurement Accuracy*
Parameter
Measurement Range
Speed
Fast
Medium
Slow
*At optimum test signal levels, frequencies, DUT values and without calibration uncertainty.
Capacitance (Cs/Cp), Inductance (Ls/Lp), Resistance (Rs/Rp), Dissipation (D) and Quality (Q)
Factors, Impedance Z , Admittance Y , Phase Angle (0-), Equivalent Series Resistance (ESR),
Conductance (Gp), Reactance (Xs), Susceptance (Bp)
Note: s = series, p = parallel, ESR equivalent to Rs