SPECIFICATIONS

Applicable Fibers

SM, MM, DSF, NZD, EDF, BIF/UBIF (Bend insensitive fiber)

Cladding Diameter

80~150 μm

Coating Diameter

100~1,000 μm

Fiber Cleave Length

5~16 mm

Average Splice Loss

SM: 0.02 dB, MM: 0.01 dB, DSF: 0.04 dB, NZD: 0.04 dB

Splice Time

7 seconds (semi-auto mode)

9 seconds (regular mode)

Heat Time

25 seconds

*6

(S922: 40 mm Sleeve, S921: 60 mm Sleeve) (Pre-heat mode)

31 seconds

*6

(S922: 40 mm Sleeve, S921: 60 mm Sleeve) (regular mode)

Splice Programs

Max. 150

Automatic Splicing Selection

SM: SM, DSF, NZD, BIF/UBIF, MM: MM

Heat Programs

Max. 18

Automatic Heating Start

Available

Applicable Sleeves

20/40/60 mm

Fiber Holding

Tight holder (Loose tube applicable) or Removal Fiber Holder System

Tension Test

1.96 N

Return Loss of Splice

60 dB or greater

Attenuation Splice Function

Intentional high splice loss of 0.1 dB to 10 dB (0.1 dB step) can be made for an inline fixed attenuator

Fiber Image Magnification

304X, 608X

Splice Memory

Max. 2,000

Image Capture Capacity

Last 100 images to be automatically captured + Up to 24 images to be stored permanently

Dimension

127W × 199D × 105H mm (not including shock absorber)

159W × 231D × 130H mm (including shock absorber)

Weight

1.9 kg (without battery)

2.3 kg (with two batteries)

Monitor

3.5” color LCD monitor

Data Output

USB ver.2.0 mini

Battery Capacity

Typical 80 splice/heat cycles with single battery

*7

Typical 200 splice/heat cycles with 2 batteries

*8

Altitude

5,000 mh

Wind Protection

Max. wind velocity of 15 m/s.

Operating Temperature

-10 to +50°C (0 to 95% Relative Humidity [Non-Condensing])

Storage Temperature

-40 to +60°C (0 to 95% Relative Humidity [Non-Condensing])

Humidity

0 to +95% RH (non-condensing)

Power Source

AC Input 100 to 240 V (50/60 Hz), DC Input 11 to 17 V without any change of hardware

*6 The first heating after turning on the power can be longer than the usual heating time.
*7 The number of the splicing and heating the machine can produce using a fully charged brand new battery at room temperature of 20°C,
semi-auto mode for splicing and regular mode for Heating. Depending on the condition of the batteries and operation environment, the number can vary.
*8 The number of the splicing and heating the machine can produce using 2 fully charged brand new batteries at room temperature of 20°C,
semi-auto mode for splicing and regular mode for Heating. Depending on the condition of the batteries and operation environment, the number can vary.

STANDARD PACKAGE

Item

P/N

Quantity

S178A Main Body

S178-A-A-0001

1

Hard Carrying Case

HCC-01

1

Battery Pack

Depending on the package

S943B

1 or 2

Spare Electrodes

S969

1pair

AC Adaptor for S178A and

S958C

S976A

1

AC Cable Cord

1

Electrode Sharpener

D5111

1

Cleaning Brush

VGC-01

1

Tool case

TCC-01

1

User Manual

FTS-B405

1