beautypg.com

Electric and magnetic field measurement, Specifi cations, Ehp-200 field analyzer – Atec Narda-EHP-200 User Manual

Page 3

background image

38

narda

Safety Test Solutions an

communications company USA • Germany • Italy

USA TEL: (1) 631 231-1700 • FAX: (1) 631 231-1711 • E-MAIL: [email protected] • www.narda-sts.us

Electric and Magnetic Field Measurement

EHP-200 Field Analyzer

Specifi cations

EHP-200 E&H FIELD ANALYSER

EHP-200 E AND H FIELD ANALYZER

RF SPECIFICATIONS

Electric Field

Magnetic Field Mode A

Magnetic Field Mode B

Auxilllary Input

Frequency Range

9 kHz to 30 MHz

9 kHz to 30 MHz

300 kHz to 30 MHz

9 kHz to 30 MHz

Measurement Range
@ 10 KHz RBW

@ Preamp ON

0.1 to 1000 V/m

0.02 to 200 V/m

0.03 to 300 A/m

6.0 mA/m to 60 A//m

3.0 mA/m to 30 A/m

0.6 mA/m to 6 A/m

-80 to 0 dBm

-94 to -14 dBm

Dynamic Range

>80 dB

Sensitivity
@ 10 kHz RBW

@ Preamp ON

0.1 V/m

0.02 V/m

30 mA/m

1 mA/m

3 mA/m

0.1 mA/m

-80 dBm

0.01 dB

Resolution

0.01 V/m

1 mA/m

0.1 mA/m

0.01 dB

Flatness

±0.5 dB (20 V/m,

from 0.1 to 27 MHz)

±0.8 dB (166 A/m,

from 0.15 to 30 MHz)

±0.8 dB (53 mA/m,

0.3 to 27 MHz)

±0.4 dB (-20 dBm)

Anisotropy

±0.8 dB at 1 MHz

Linearity

0.5 dB @ 1 MHz from Full Scale to -60 dB Full Scale

Typical Accuracy at 1 MHz

±0.8 dB @ 20 V/m

±0.8 dB @ 53 mA/m

±0.8 dB @ 53 mA/m

±0.3 dB@-10 dBm

Maximum Frequency Span

6 kHz to 30 MHz

Resolution Bandwidths Available

1 kHz, 3 kHz, 10 kHz, 30 kHz, 100 kHz, 300 kHz

Rejection to E-field

> 20 dB

Rejection to H-field

> 20 dB

Calibration Errors

Stored in internal EEPROM

Temperature Error

0.02 dB/

o

C

GENERAL SPECIFICATIONS

Preamplifier

Selectable ON/OFF, 14 dB gain

Reading Units

V/m, A/m, mT, mW/cm

2

, W/m

2

Optical Link

Maximum length of 80 m

Internal Battery

3.7 V, 3.6 Ah, Li-ion, rechargeable

Battery Operation Time

> 8 hours (recharging time approximately 8 hours)

External Supply

10 – 15 VDC, 500 mA

Firmware Update

Via Optical Fiber

Operating Temperature

-10

o

C to +50

o

C

Storage Temperature

-20

o

C to +70

o

C

Dimensions and Weight

3.6 x 3.6 x 4.3 inches (92 x 92 x 109 mm), 1.22 lb (550 g)

ORDERING INFORMATION

Part Number

EHP-200

Accessories Supplied

10 meter Fiber Optic Cable ( FO-8053/10), Fiber Optic Converter (FO to RS-232, 8053-OC), Soft Carrying

Bag ( 8053-SC), Battery Charger ( 8053-BC), 50 cm Plastic Pole and Tripod, PC Software, Operating Manual,

Calibration Certificate

Optional Accessories

EHP-200 Palm Kit

www.narda-sts.com