SPECIFICATIONS
a
Probe EA … ED5091
Electric (E-)Field
Frequency range
(b)
300 kHz to 50 GHz
(EA5091, EC5091, ED5091)
3 MHz to 50 GHz
(EB5091)
Type of frequency response
Shaped, see table on page 2
Measurement range
0.5 to 600 % of Standard (Power Density)
Dynamic range
30 dB
CW damage level
2000 % of Standard
700 mW/cm²
Peak damage level
(c)
32 dB above Standard
Sensor type
Combined diodes/ thermocouples
Directivity
Isotropic (Tri-axial)
Readout mode / spatial assessment
Combined 3-axis (RSS)
UNCERTAINTY
Flatness of frequency response
(d)
Calibration uncertainty not included
±2 dB from Standard
Calibration uncertainty
(e)
±1.5 dB (< 1.8 GHz)
±0.7 dB (1.8 GHz to 40 GHz)
±1 dB (@ 45.5 GHz)
Linearity
Referred to 100 %
±3 dB (< 4 % of Standard)
±1 dB (4% to 12 % of Standard)
±0.5 dB (12 % to 600 % of Standard)
Isotropic response
(f)
typ. ±1 dB (≥ 10 MHz)
GENERAL SPECIFICATIONS
Calibration frequencies
0.3/ 3/ 10/ 30/ 100/ 300/ 750 MHz 1/ 1.8/ 2.45/ 4/ 8.2/ 10/ 18/ 26.5/ 40/ 45.5 GHz
Recommended calibration interval
12 months
Temperature range
Operating
Non-operating (transport)
0 °C to +50 °C
-40 °C to +70 °C
Humidity
5 to 95 % RH @ ≤25 °C
≤23 g/m³ absolute humidity
Size
350 mm x 104 mm Ø
Weight
240 g
Compatibility NBM-500
series
meters
(a)
Unless otherwise noted specifications apply at reference condition: device in far-field of source, ambient temperature 23±3 °C, relative air humidity 40% to 60%, sinusoidal signal
(b)
Cutoff frequency at approx. -3 dB
(c)
Pulse length 1µsec, duty cycle 1:1000
(d)
Frequency response can be compensated for by the use of correction factors stored in the probe memory
(e)
Accuracy of the fields generated to calibrate the probes
(f)
Uncertainty due to varying polarization (verified by type approval test for meter with probe). Ellipse ratio included and calibrated for each probe
ORDERING INFORMATION
Part number P/N
Probe EA5091, Shaped E-field, FCC for NBM, 300 kHz - 50 GHz, isotropic
2402/07
Probe EB5091, Shaped E-field, IEEE for NBM, 3 MHz - 50 GHz, isotropic
2402/08
Probe EC5091, Shaped E-field, SC6 Canada for NBM, 300 kHz - 50 GHz, isotropic
2402/09
Probe ED5091, Shaped E-field, ICNIRP for NBM, 300 kHz - 50 GHz, isotropic
2402/10