APPLICATIONS
Monitoring RF signals (occupational limits)
Mobile, telecom, broadcast and industrial applications are covered by
the standard version of RadMan and RadMan XT (with data logger). The
E field and H field sensors are shaped to match the higher level of two-
tier standards, i.e. the “Controlled”, “Occupational” or “RF Worker” limits.
Monitoring pulsed radar signals
The “Fast Response” versions of RadMan and RadMan XT are
designed to detect the peaks of sharp and narrow radar pulses. The
ICNIRP standard, for example, requires peak detection when the ratio of
peak to average power is greater than 30 dB. Peak detection is obtained
by reducing the 1 second integration time for E-fields down to 30
milliseconds for all “Fast” models.
Monitoring workplaces
Fixed on a non-conductive tripod (optional accessory) the RadMan can
be used for workplace radiation monitoring. This application is ideal for
areas to be monitored occasionally. For permanent surveillance Narda
offers the wall mountable Smarts II
TM
.
Searching for leaks
As a locating unit all RadMan and RadMan XT versions can be used to
locate leaks on waveguides and coaxial screw connectors (picture on
the right). A non-conductive extension with handle is available as an
optional accessory.