beautypg.com

Applications – Atec Narda-RadMan User Manual

Page 3

background image


NSTS 1013-E0101M

3 / 5

Subject to change

APPLICATIONS

Monitoring RF signals (occupational limits)

Mobile, telecom, broadcast and industrial applications are covered by

the standard version of RadMan and RadMan XT (with data logger). The

E field and H field sensors are shaped to match the higher level of two-
tier standards, i.e. the “Controlled”, “Occupational” or “RF Worker” limits.

Monitoring pulsed radar signals
The “Fast Response” versions of RadMan and RadMan XT are
designed to detect the peaks of sharp and narrow radar pulses. The

ICNIRP standard, for example, requires peak detection when the ratio of

peak to average power is greater than 30 dB. Peak detection is obtained

by reducing the 1 second integration time for E-fields down to 30
milliseconds for all “Fast” models.

Monitoring workplaces

Fixed on a non-conductive tripod (optional accessory) the RadMan can

be used for workplace radiation monitoring. This application is ideal for

areas to be monitored occasionally. For permanent surveillance Narda

offers the wall mountable Smarts II

TM

.

Searching for leaks

As a locating unit all RadMan and RadMan XT versions can be used to

locate leaks on waveguides and coaxial screw connectors (picture on

the right). A non-conductive extension with handle is available as an

optional accessory.

RadMan XT used as an instrument for

leak detection

RadMan XT used as a personal

monitor for radio frequency fields