Spectrum analyzer – Atec Anritsu-MS2723B User Manual
Page 2
Page 2 of 20
Spectrum Master™ MS2723B Spectrum Analyzer Specifications
Spectrum Analyzer
Measurements
Smart Measurements
Field Strength (uses antenna calibration tables to measure dBm/m
2
or dBmV/m)
Occupied Bandwidth (measures 99% to 1% power channel of a signal)
Channel Power (measures the total power in a specified bandwidth)
ACPR (adjacent channel power ratio)
AM/FM/SSB Demodulation (wide/narrow FM, upper/lower SSB), (audio out only)
C/I (carrier-to-interference ratio)
Emission Mask (recall limit lines as emission mask)
Setup Parameters
Frequency
Center/Start/Stop, Span, Frequency Step, Signal Standard, Channel #
Amplitude
Reference Level (RL), Scale, Attenuation Auto/Level, RL Offset, Pre-Amp On/Off, Detection
Span
Span, Span Up/Down (1-2-5), Full Span, Zero Span, Last Span
Bandwidth
RBW, Auto RBW, VBW, Auto VBW, RBW/VBW, Span/RBW
File
Save, Recall, Delete, Directory Management
Save/Recall
Setups, Measurements, Limit Lines, Screen Shots Jpeg (save only), Save-on-Event
Save-on-Event
Crossing Limit Line, Sweep Complete, Save-then-Stop, Clear All
Delete
Selected File, All Measurements, All Mode Files, All Content
Directory Management
Sort Method (Name/Type/Date), Ascend/Descend, Internal/USB CF, Copy
Application Options
Impedance (50 Ω, 75 Ω, Other)
Sweep Functions
Sweep
Single/Continuous, Manual Trigger, Reset, Detection, Minimum Sweep Time, Trigger Type,
Gated Sweep (see Option 0090)
Detection
Peak, RMS, Negative, Sample, Quasi-peak
Triggers
Free Run, External, Video, Change Position, Manual
Trace Functions
Traces
Up to three Traces (A, B, C), View/Blank, Write/Hold, Trace A/B/C Operations
Trace A Operations
Normal, Max Hold, Min Hold, Average, # of Averages, (always the live trace)
Trace B Operations
A B, BC, Max Hold, Min Hold
Trace C Operations
A C, BC, Max Hold, Min Hold, A – B C, B – A C, Relative Reference (dB), Scale
Marker Functions
Markers
Markers 1-6 each with a Delta Marker, or Marker 1 Reference with Six Delta Markers,
Marker Table (On/Off), All Markers Off
Marker Types
Style (Fixed/Tracking), Noise Marker, Frequency Counter Marker
Marker Auto-Position
Peak Search, Next Peak (Right/Left), Peak Threshold %, Set Marker to Channel,
Marker Frequency to Center, Delta Marker to Span, Marker to Reference Level
Marker Table
1-6 markers frequency and amplitude plus delta markers frequency offset and amplitude
Limit Line Functions
Limit Lines
Upper/Lower, On/Off, Edit, Move, Envelope, Advanced, Limit Alarm, Default Limit
Limit Line Edit
Frequency, Amplitude, Add Point, Add Vertical, Delete Point, Next Point Left/Right
Limit Line Move
To Current Center Frequency, By dB or Hz, To Marker 1, Offset from Marker 1
Limit Line Envelope
Create Envelope, Update Amplitude, Number of Points (41), Offset, Shape Square/Slope
Limit Line Advanced
Type (Absolute/Relative), Mirror, Save/Recall
Frequency
Frequency Range
9 kHz to 13 GHz (usable to 0 Hz), Preamp 100 kHz to 4 GHz
Tuning Resolution
1 Hz
Frequency Reference
Aging: ± 1.0 ppm/10 years
Accuracy: ± 0.3 ppm (25 °C ±25 °C) + aging
Frequency Span
10 Hz to 13 GHz including zero span
Span Accuracy
Same as frequency reference accuracy
Sweep Time
10 µs to 600 seconds in zero span, autoset in non-zero span
Sweep Time Accuracy
± 2% in zero span
Bandwidth
Resolution Bandwidth (RBW)
1 Hz to 3 MHz in 1–3 sequence ± 10% (1 MHz max in zero-span) (–3 dB bandwidth)
Video Bandwidth (VBW)
1 Hz to 3 MHz in 1–3 sequence (–3 dB bandwidth)
RBW with Quasi-Peak Detection
200 Hz, 9 KHz, 120 kHz (-6 dB bandwidth)
VBW with Quasi-Peak Detection
Auto VBW is On, RBW/VBW = 1