Atec Commtest-VB7 User Manual
Page 2
Contact us on +64 3 943 0700, [email protected] or visit www.commtest.com
SPECIFICATIONS
vb7
®
DATA ANALYZERS
REMARKS
Sensor Input
Sensor input
2 channels
Simultaneous sampling
Compatible sensor types
Accelerometer, velocity,
displacement, current
AC coupled range
16 V peak-peak
Allows for ± 8 V sensor output
swing (± 80 g)
DC coupled ranges
0 V to 20 V, -10 V to 10 V,
-20 V to 0 V
E.g. for reading prox-probe gap
Connectors
2 x BNC (CH1/CH2)
Safety feature: Break-free
inline connector
Analog to digital conversion
24-bit ADC
Sensor excitation current
0 mA or 2.2 mA (configurable),
24 V maximum
2.2 mA required for ICP®-type
accelerometer
Sensor detection
Warns if short circuit or not
connected
Tachometer
Sensor
Laser sensor with reflective
tape
Sensor triggers on beam
reflection
Laser sensor range
10 cm to 2 m nominal
Depends on size of reflective
tape
Other Sensor types supported
Contact, TTL pulse,
Keyphasor
®
Optically isolated input
Power supply to sensor
5 V, 50 mA
TTL pulse rating
3.5 V (4 mA) min, 28 V (5 mA)
max, off-state 0.8 V
Keyphasor® thresholds
7.7 ± 0.5 V, 13.3 ± 0.8 V,
18.8 ± 1 V
Nominally 8 V, 13 V, 18 V
Speed range
10 to 300 000 RPM (0.2 to
5 kHz)
Pulse width at least 0.1 ms
Accuracy
+/- 0.1%
Output to drive strobe
Up to 140 Hz (8400 CPM)
Typical. Depends on strobe
type. Special cable required.
Parameter Indication
Maximum levels
> 1000 g (10 000 m/s2)
> 1000 in/sec (25 000 mm/s)
> 100 in (2500 mm),
> 10 000 amps
Effective limit is sensor
sensitivity and output voltage
Dynamic signal range
> 95 dB (typical at 400 line
resolution)
Harmonic distortion
Less than -70 dB typical
Other distortions and noise
are lower
Units
g or m/s2
in/s or mm/s
mil or mm or µm
adB, vdB, amps, user-defined
0-peak, peak-peak or RMS.
Auto-scale by 1000x when
required
US & SI options for both adB
& vdB
Magnitude & cursors
Overall RMS value, dual
cursors, harmonics
Digital readouts on chart
Base accuracy
± 1% (approx. 0.1 dB)
For DC level: % of full scale.
For AC signal: % of reading
High frequency attenuation
≤ 0.1 dB 100 Hz to 10 kHz
≤ 3 dB >10 kHz to 40 kHz
Attenuation tolerances are in
addition to base accuracy
AC coupling attenuation
≤ 0.1 dB 10 Hz to <100 Hz
≤ 3 dB 1 Hz to <10Hz
Attenuation due to Integration
≤ 0.1 dB 1 Hz to <100 Hz
≤ 1.5 dB 0.2 Hz to <1 Hz
≤ 0.1 dB 10 Hz to <100 Hz
≤ 1.5 dB 1 Hz to <10 Hz
Low freq. mode: When coupling
= DC, Fmax
≤ 100 Hz.
Normal mode: Applicable in all
other cases.
Values apply to single
integration (accel. to veloc.).
Double the values for double
integration (accel. to displ.).
Spectrum Display
Fmax ranges
25, 50, 100, 125, 150, 200,
300, 400, 500, 600, 800, 1000,
1200, 1600, 2000, 2500, 3000,
4000, 5000, 6000, 8000, 10 000,
15 000, 20 000, 30 000,
40 000 Hz
Or equivalent CPM values
Or orders-based from 1X to
999X
Fmin possible range
0 to Fmax
Instrument zeroes all spectral
lines below Fmin
Resolution
400, 800, 1600, 3200, 6400 lines
3200 lines max. for dual
channel measurements
Frequency scale
Hz, CPM, Orders
Linear scale with zooming
Amplitude scale
Acceleration, velocity,
displacement or current
Linear or log scales, auto or
manual scaling
Window shapes
Hanning, rectangular
Overlap
(0, 12.5, 25, 37.5, 50, 62.5, 75,
87.5) %
Dependent on Fmax and
number of lines
Number of averages
1, 2, 4, 8, 16, 32, 64, 128
Increases sampling time
proportionally
Averaging types
Linear, exponential,
peak hold, synchronous
SPECIFICATIONS
vb7
®
DATA ANALYZERS
REMARKS
Spectrum Display Continued
Demodulation bandwidths
23 bandwidth options
From 125 Hz to 1250 Hz up to
16 kHz to 20 kHz
6Pack
Up to 40 kHz & 3200 lines 1
channel
Up to 20 kHz & 1600 lines 2
channel
Spectrum and waveform for
low freq, high freq, demod.
Order tracking
Up to 6 kHz Fmax, Orders-based
Tachometer required, mounted
on high-speed shaft
Order tracking - Distortion
Less than -65 dB
Within 50% to 200% speed
variation during recording
Waveform Display
Number of samples
1024, 2048, 4096, 8192, 16 384
Time scale
10 ms to 256 seconds
Or orders based from 1 to
999 revs
Time synchronous averages
1, 2, 4, 8, 16, 32, 64, 128
Only available when
tachometer triggered
Long time waveform Fmax
25 Hz to 40 kHz
20 kHz dual channel
Long time waveform duration
14.7 million samples (total over
channels)
E.g. for Fmax 1 kHz, Fsample =
2.56 kHz and Duration= 1.6 hrs
Logging & Analysis
Output formats
LCD screen, Ascent, XML
Data storage
Dual 1 GB non-volatile flash
memories
Database mirror copy on
second flash memory
Data storage structure
Folders / machines / points /
locations / routes
No limits are applied, 50
character names
Max Folder size
10 000 measurement locations
Balancing
Planes
Up to 2 planes, 2 sensors
Speed range
30 to 60 000 RPM
Measurement type
Acceleration, velocity,
displacement
Weight modes
Angle 0° to 360°, fixed position,
circumference arc
E.g. weights on fan
blades, linear dist. around
circumference
Remove trial weights
Yes, No
Automatic recalculation
Manual data entry
Yes
Allows re-entry of previous
balance jobs
Storage
Against machines in data
structure
No limits applied
Display & Communication
Display
Graphic Grayscale LCD
White LED Backlight
Resolution & size
480 x 320 (HVGA), 5.5” (140
mm)
Readable in direct sunlight
Supported Languages
ENG, FRE, SPA, POR, RUS, CHI
Firmware releases in English,
Communication with PC
USB and Ethernet
PROFLASH allows instrument
software to be upgraded
USB host port
USB 2.0, supplying 5V, 250mA
Save folders to USB flash drive
Battery & Charger
Battery type
Custom Lithium Ion pack,
7.4 V, 4500 mAh
Operating time
10 hours
Backlight on (60 second
time-out)
Charger type
Internal charging,
automatic control
External power pack 12 V DC,
3 A output
Charge rate
3 A nominal
3 hours for complete charge
Mechanical
Size
9.9” W x 5.8” L x 2.4” H
(252 x 148x 60) mm
Weight
2.7 lb (1.2 kg)
Including battery and strap
Environmental
Operating temperature
14 °F to 122 °F (-10 to 50) °C
Storage temperature &
humidity
-4 °F to 140 °F (-20 to 60) °C,
95% RH
If storage exceeds 1 month: Up
to 95 F (35 C), 85% RH
EMC
EN61326
Ruggedness
4’ (1.2 m) drop onto concrete,
IP65
Procedure: 26 drops following
MIL-STD-810F-516.5-IV