System Characteristics
Configuration Times (typical)
9
Function change
Standard
100
ms
Pulse
660
ms
Built-in arb
220 ms
Frequency change
20 ms
Amplitude change
50 ms
Offset change
50 ms
Select user arb
< 900 ms for < 16K pts.
Modulation change < 200 ms
Arb Download Times GPIB/RS-232 (115Kbps)
Arb Length Binary ASCII Integer ASCII Real
64K points 48 sec 112 sec
186 sec
16K points 12 sec 28 sec
44 sec
8K points
6 sec
14 sec
22 sec
4K points
3 sec
7 sec
11 sec
2K points
1.5 sec 3.5 sec
5.5 sec
Trigger Characteristics
Trigger input
Input level
TTL compatible
Slope
rising or falling,
(selectable)
Pulse width
> 100 ns
Input impedance
10 kΩ, DC coupled
Latency
Burst
< 100 ns (typical)
Sweep
< 10 µs (typical)
Jitter (rms)
Burst
1 ns; except pulse,
300 ps
Sweep
2.5
µs
Trigger output
Level
TTL compatible into
50Ω
Pulse width
> 450 ns
Maximum rate
1 MHz
Fanout
≤ 4 Agilent 33250A’s
(or
equivalent)
Clock Reference
Phase Offset
Range
-360° to +360°
Resolution 0.001°
External Reference Input
Lock range
10 MHz ± 35 kHz
Level
100 mVpp to 5 Vpp
Impedance
1 kΩ nominal, ac
coupled
Lock time
< 2 s
Internal Reference Output
Frequency 10
MHz
Level
632 mVpp (0 dbm),
nominal
Impedance
50Ω nominal, ac
coupled
Sync Output
Level
TTL compatible
into > 1 kΩ
Impedance
50 Ω nominal
General
Power supply
100-240 V, 50-60 Hz
100-127 V, 50-400 Hz
Power consumption 140 VA
Operating temp.
0°C to 55°C
Storage temp.
-30°C to 70°C
Stored states
4 named user configu-
rations
Power on state
default or last
Interface IEEE-488
and
RS-232 std.
Language SCPI-1997,
IEEE-488.2
Dimensions (WxHxD)
Bench top
254 x 104 x 374 mm
Rackmount
213 x 89 x 348 mm
Weight 4.6
kg
Safety designed to
EN61010-1, CSA1010.1,
UL-311-1
EMC tested to
IEC-61326-1
IEC-61000-4-3 criteria B
IEC-61000-4-6 criteria B
Vibration and shock MIL-T-28800E, Type III,
Class 5
Acoustic noise
40 dBA
Warm-up time
1 hour
Calibration interval
1 year
Warranty 1
year
3
1 Harmonic distortion at low amplitudes is limited by a -70 dBm floor
2 Harmonic distortion at 40 MHz only is -33 dBc
3 Spurious noise at low amplitudes is limited by a -75 dBm floor
4 Edge time decreased at higher frequency, 3.5 nS (typical)
5 20 mVpp to 20 Vpp into open-circuit load
6 dB rounded to 1 digit, instrument adheres to “%” specification
7 Short-circuit protected to ground at all times
8 Sine and square waveforms above 25 MHz only with infinite burst count
9 Time to change parameter and output new signal