beautypg.com

Specifications and characteristics (continued), Power meter functions – Atec Agilent-N432A User Manual

Page 7

background image

7

Power meter functions

Accessed by key entry

Either hard keys, or soft key menu, and programmable

Zero

Zeros the meter. (Power reference calibrator is switched off during zeroing)

Frequency

Entered frequency range is used to interpolate the calibration factors table. Frequency range from
1 kHz to 999.9 GHz

Bridge resistance type

Setting the bridge resistance, user-selectable Rmeasure (factory set) or Ruser (measured by user)

Cal factor

Sets the cal factor versus frequency for the calibration factor for power sensor. Range: 1.00% to
150.00%, in 0.01% increments. Cal factor can be set from either CF table or single entry CF

Relative

Displays all successive measurements relative to the last displayed value

Offset

Allows power measurements to be offset by –100 to +100 dB, settable in 0.001dB increments, to
compensate for external loss or gain. Offset table can be set from either CF table or single entry
CF

Save/recall

Store up to 10 instrument states via the save/recall menu

Measurement averaging

Selectable from 4 to 128

Duty cycle

Duty cycle values between 0.001 to 99.999%, in 0.001% increments, can be entered to display a
peak power representation of measured power. The following equation is used to calculate the
displayed peak power value: peak power = measured power/duty cycle

Limits

High and low limits can be set in the range –150.000 to +230.000 dBm, in 0.001 dBm increments

Preset default values

dBm mode, relative off, power reference off, duty cycle off, offset off, frequency 50 MHz,
measurement averaging 16, free run

Display

Color display with selectable single and split screen formats are available. User selectable on
digital measurement type or analog scale presentation

1 mW (0 dBm) power reference

Accuracy (for two years)

SWR

1.0 mW (0.0 dBm), 50 MHz from type N (f) connector on the front panel, for power meter/sensor
function check.

± 0.4% (25 ± 10 ºC)
± 1.2% (0 to 45 ºC)

1.06 maximum