Measurement Characteristics

Measurement Techniques

Stat Mode (4500A only): Continuous sampling 0.5 M Samples per

second

Power Mode:

Random repetitive sampling system that
provides pre- and post-trigger data

Maximum Sample Rate

1MHz

Memory Depth

4 K/Channel

Vertical Resolution

0.025%, 12-bit A/D converter

Waveform Averaging

1 to 10,000 samples per data point

Waveform Storage

Two reference waveforms in internal non-
volatile memory

Trigger Channel Bandwidth: > 30 MHz typical

Sensor Characteristics

Power Detection Technique: Dual diode with selectable detector bandwidth
Log Amplifier:

The logarithmic amplifier in the sensor enables
the instrument to measure and analyze
changes in power exceeding 60 dB in a single
display range.

Internal Data

Sensor calibration factors, frequency range,
power range, sensor type, serial number and
other sensor dependent information are stored
in EEPROM within the peak power sensor.

Sensor Cable

The sensor cable is detachable from both the
sensor and instrument. The standard cable
length is 5 feet. Other cable lengths are 10 feet,
20 feet, 25 feet and 50 feet.

Rear Panel Connections

External Calibrator
Pulse Input

Provides a means of applying an external TTL
level signal to control the pulse rate and duty
cycle of the calibrator output (50 ohm input
impedance)

IEEE-488 Interface

Complies with IEEE-488-1978. Implements
AH1, SH1, T6, LE0, SR1, RL1, PP0, DC1, C0,
and E1

RS-232 Interface 1

Serial Printer/Plotter interface

RS-232 Interface 2

Diagnostic interface

Parallel Port

Parallel Pinter/Plotter interface

Optional Connectors

5

Rear Panel
Channel 1 and 2, Trigger 1 and 2, calibrator
output

Physical and Environmental

General

Manufactured to the intent of MIL-T-28800E,
Type III, Class 5, Style E

Disk Drive

3.5-inch, 1.44-MB (DOS compatible)

Display

VGA-compatible, seven-inch diagonal color
CRT with 640 x 480-pixel resolution. Waveform
display area resolution is 501 x 281.

Operating Temperature

0 to 50

O

C

Storage Temperature

-40 to 75

O

C

Humidity

95% + 5% maximum (non-condensing)

Altitude

Operating: 10,000 feet (3,000 meters)

Power Requirements

90 to 260 VAC, 47 to 440 Hz, 200 VA maximum

Dimensions

17.25 inches (43.8 cm) wide, 7 inches (17.8 cm)
high, 22 inches (55.9 cm) deep

Weight

38 pounds (17.2 kg) with second channel
installed

Hard Copy Output

The screen can be output to a printer or plotter
on the RS-232m parallel, IEEE-488 devices, or
to a file on disk.

HPGL Plotters

HP7475, HP7470, ATT 435

Printers

ThinkJet, LaserJet II

Sensors

See Appendix D or the Boonton Electronics Sensor Data Manual for detailed
specifications of Boonton Peak Power Sensors.

CE Mark

Boonton Electronics Corporation declares conformity of the Model 4400A
RF Peak Power Meter and the Model 4500A RF Peak Power
Meter/Analyzer to the following European Community Council (ECC)
regulations:

Directives: 89/336/EEC/93/68/EEC and 73/23/EEC/93/68/EEC

Standards: EN55011, EN50082-1, EN61010-1

Notes

1

Sensor dependent

2

Sensitives are decreased by a factor of two in the split-screen mode.

3

CW mode

4

CW mode, 0 to 40

O

C

5

Available with optional second channel

ADDITIONAL PRODUCTS AND SERVICES
AVAILABLE* FROM BOONTON ELECTRONICS
CORPORATION

Model 4530 series RF Peak Power Meter 10 kHz to 40 GHz
Model 4230A series RF Power Meter 10 kHz to 40 GHz
Model 9230 series RF Voltmeter 200

m to 300 V, 10 Hz to 2.5 GHz

Model 5230 series Universal RF Power Meter/Voltmeter 200

mV

to 300 V, 10 Hz to 100 GHz
Model 4300 RF Power Meter, multi-channel 10 kHz to 40 GHz
Model 92EA RF Millivoltmeter 200

mV to 300 V, 10 Hz to 1.2 GHz

Model 7200 Capacitance Meter 0 to 2,000 pF, 1 MHz, Analog

display
Model 72B Capacitance Meter 1 pF to 3,000 pF, 1 MHz, Analog
display

Model 8201 Modulation Analyzer 100 kHz to 2.5 GHz
Model 8701 VXI Modulation Analyzer 100 kHz to 2.5 GHz
Model 1121 Audio Analyzer 10 Hz to 140 kHz

Service and Support You Can Count On

Boonton Electronics backs all of its products with a
full range of test, repair, upgrade and calibration
services assuring that all your instrumentation
remains accurate, reliable and conforms to original
factory specifications.
Services include:
· Certified Repairs (NIST Traceable)

· Repair Warranty (Six Months, Materials and

Labor)
· Automatic Instrument/Software Upgrades*

· 10-day Turnaround

· Priority Service

· Flexible Service, Repair, and Calibration

Contracts
Contact our Customer Service Department at
(973) 386-9696 for details and pricing information.