Atec Chroma-6310A Series User Manual
Page 4
panel desCription
1
4
3
2
5
6
7
8
14
10
11
15
20
25
16
19
23
21
22
24
18
17
13
12
9
Load Module
Mainframe Controller
timing funCtion
The 6310A series of loads include a unique timing & measurement function, which allows
precise time measurements in the range of 1ms to 86,400s. This feature allows the user
to set the final voltage & timeout values for battery discharge testing and other similar
applications.
For example, the figure on the right shows the 6310A internal timer starting at load ON, and
ending when the battery voltage reaches the final voltage.
The Timing function can be used in testing battery and super capacitor discharge, or other
similar applications.
Final Voltage
Time
Battery Discharge Testing
Battery Voltage
Loading Current
26
digital i/o
The digital I/O interface makes the 6310A DC
Load the ideal choice for automated testing
requirements. Through the digital I/O, the
6310A can accept digital signals to trigger its
functions (Load On/Off, OCP test, etc.) as well
as current output status signals.
Pin 1
Reserved
Pin 2
DGND
Pin 3
DGND
Pin 4
DGND
Pin 5
DGND
Pin 6
Load ON/OFF (O/P)
Pin 7
Total Pass (O/P)
Pin 8
Total Fail (O/P)
Pin 9
Short Signal (O/P)
Pin 10 Protection Signal (O/P)
Pin 11
External Load ON/OFF (I/P)
Pin 12 Reserved
Pin 13 Reserved
Pin 14 DGND
Pin 15 External Trig. For
Sequences Run (I/P)
Pin
Definition