background image

4

Agilent 8719ES and 8720ES
with 3.5 mm test ports

Standard, Options 400, 012

1

, 089, or any combina-

tion of these options.

Calibration kit: 85052B, 3.5-mm with sliding loads
Cables: HP 85131F 3.5-mm flexible cable set
IF bandwidth: 10 Hz
Averaging: None (8 during isolation calibration)

System dynamic range

0.05 to 0.84 GHz

77 dB

0.84 to 20 GHz

100 dB

System performance

Corrected measurement port specifications

Frequency range (GHz)

0.05 to 0.5

0.5 to 2

2 to 8

8 to 20

Directivity

48 dB

48 dB

44 dB

44 dB

Source match

40 dB

40 dB

33 dB

31 dB

Load match

48 dB

48 dB

44 dB

44 dB

Reflection tracking

±(0.006 dB + 0.02 dB/°C)

±(0.006 dB + 0.03 dB/°C)

±(0.006 dB + 0.03 dB/°C)

±(0.008 dB + 0.04 dB/°C)

Transmission tracking

±(0.017 dB + 0.02 dB/°C)

±(0.018 dB + 0.03 dB/°C)

±(0.066 dB + 0.03 dB/°C)

±(0.099 dB + 0.04 dB/°C)

Maximum output power
+5 dBm

Magnitude

Phase

Magnitude

Phase

1. Internal test set not bypassed.

Measurement uncertainty

Reflection measurements

Transmission measurements