Atec EM-Test-CSS500N10 User Manual
Page 2
DATA SHEET > CSS 500N10 > 20130626
TECHNICAL DETAILS
CSS 500N10 SURGE CURRENT
GENERATOR
WAVE SHAPE 8US/20US
Output voltage
(open circuit)
250V to 6,000V +/-10%,
Step width 50V
Output current
(short circuit)
0.4kA to 10kA +/-10%
Front time
8us +/-20%
Time to half value
20us +/-20%
Undershoot
Max. 20%
TRIGGER
Trigger of events
Automatic, manual, external
External
External pulse release
Repetition rate
Max. 10 pulses per minute,
depending on the charging voltage
Counter
1 to 30,000 or endless
OUTPUT
Direct
Via high current connector
Optional: Test box
Polarity
Positive, negative, alternating
MEASUREMENTS
CRO I-monitor
10Vp at 10kA
CRO V-monitor
6Vp at 6kV
Peak current
10kA in the LCD display
Peak voltage
6kV in the LCD display
TEST ROUTINES
Quick Start
Immediate start; easy-to-use and
fast
User Test routines
Change polarityafter n pulses
Change charging voltage after n
pulses
Customer specific test routines
Service
Service, set-up, self test
INTERFACE
Optical interface
Opto link, 3 m cable
USB A connector
Parallel interface
IEEE 488, addresses 1 to 30
SAFETY
Safety circuit
Control input (24Vdc)
Warning lamp
Floating output contact
Safety interlock
Security switch off (with test box)
GENERAL DATA
Dimensions,
weight
19"/6HU,
approx. 30kg
Supply voltage
115/230V +10/-15%
Fuses
2 x T1AT (230V) or 2 x T2AT (115V)
www.emtest.com