Atec EM-Test-TSS500M4 User Manual
Page 3
DATA SHEET > TSS 500M4 > 20080219
TECHNICAL DETAILS
AS PER ITU AND ETS RECOMMENDATIONS
Pulse 1.2/50us
Voltage (o.c.)
160V - 4,000V ±10%
Rise time*)
1.0us ± 30%
Pulse duration*)
50us ± 20%
Energy storage
capacitor
1uF
Pulse 10/700us
Rise time*)
6.5us ± 30%
Pulse duration*)
700us ± 20%
Energy storage
capacitor
20uF
Polarity
Positive, negative or alternating
Counter
1 - 30,000 or endless
*) definition of waveform parameters
as per IEC 469-1. As per IEC
61000-4-5 this is considered to be
equal to the waveform parameter
definition as per IEC 60-1 for the
1.2/50us pulse and CCITT for the
10/700us pulse.
AS PER FCC PART 68, PULSE B
Voltage (o.c.)
160V - 4,000V ±10%
Front time
9us ± 30%
Decay time
720us ± 20%
Current (s.c.)
4A - 100A
Front time
5us ± 30%
Decay time
320us ± 20%
Energy storage
capacitor
20uF
Polarity
Positive, negative or alternating
Counter
1 - 30,000 or endless
AS PER IEC 61000-4-5
Pulse 10/700us
Voltage (o.c.)
160V - 4,000V ±10%
Rise time*)
6.5us ± 30%
Pulse duration*)
700us ± 20%
Current (s.c.)
4A - 100A
Rise time*)
4us ± 20%
Pulse duration*)
300us ± 20%
Energy storage
capacitor
20uF
Polarity
Positive, negative or alternating
Counter
1 - 30,000 or endless
*) definition of waveform parameters
as per IEC 469-1. As per IEC
61000-4-5 this is considered to be
equal to the waveform parameter
definition as per IEC 60-1 for the
1.2/50us pulse and CCITT for the
10/700us pulse.
COUPLING AS PER
ITU-T
2-wire: T1 and T2 with 25ohm each
4-wire: T1,T2,T3,T4 with 100ohm
each
FCC part 68
2 wire: T1 and T2 with 25ohm each
IEC 61000-4-5
External networks are required
(options)
TRIGGER
Automatic
Automatic pulse release
Manual
Single pulse release
External
External pulse release
CRO trigger
5V trigger signal for oscilloscope
TEST ROUTINES
Quick Start
Immediate start; easy to use and fast
User Test routines
Change Polarity after n pulses
Change voltage after n pulses
Standard Test
routines
IEC 61000-4-5 Level 1,000V
IEC 61000-4-5 Level 2,000V
IEC 61000-4-5 Level 4,000V
FCC part 68, Pulse B Metallic
1,000V
FCC part 68, Pulse B Longitudinal
1,500V
Service
Service, setup, self test