DATA SHEET > OCS 500N6 > 20111021

TECHNICAL DETAILS

OCS 500N6 MODELS

OCS 500N6.5

With built-in CDN 250V / 16A

OCS 500N6.6

With built-in CDN 250V / 32A

OCS 500N6.7

With built-in CDN 3x440V / 16A

OCS 500N6.8

With built-in CDN 3x440V / 32A

RINGWAVE AS PER EN/IEC 61000-4-12 AND ANSI/IEEE
C62.41

Voltage (o.c.)

250V - 6,000V ± 10%

Rise time

0.5us ± 30%

Oscillation
frequency

100kHz ± 10%

Decaying

Ratio of peak 2 to peak 1: 0.4 - 1.1
Ratio of peak 3 to peak 2: 0.4 - 0.8
Ratio of peak 4 to peak 3: 0.4 - 0.8

Source impedance

12ohm and 30ohm ± 20%
200ohm for I/O lines (CNV 504M)

Peak current (s.c.)

Max. 500A @ 12ohm or
Max. 200A @ 30ohm internally;
Max. 30A @ 200ohm, with CNV
504M

Rise time

< 1us

Oscillation
frequency

100kHz ± 10%

Polarity

Positive, negative

Repetition rate

1 to 60 transients per minute

SLOW DAMPED OSCILLATORY WAVES AS PER EN/IEC
61000-4-18

Voltage (o.c.) at
HV output

250V - 3,000V ± 10%

Voltage (o.c) at
line output

250V - 2,500V ± 10%

Rise time

75ns ± 20%

Oscillation
frequency

100kHz and 1MHz ± 10%

Decaying

Peak 5 to be > 50% of peak 1 value
Peak 10 to be < 50% of peak 1 value

Source impedance

200ohm ± 20%

Polarity

Positive, negative

Repetition rate

Max. 50/s for 100kHz and
Max. 500/s for 1MHz

Burst duration

At least 2s

TRIGGER CIRCUIT

Release of pulses

Automatic, manual, external

Synchronization

0° - 360°, resolution 1°

OUTPUT

Direct

Via HV-safety lab connectors

Coupling mode

Line to line
Line(s) to ground (PE)

DUT supply
1ph-16A

AC: 250V/16A; 50/60Hz
DC: 250V/10A

DUT supply
1ph-32A

AC: 250V/32A; 50/60Hz
DC: 250V/20A

DUT supply
3ph-16A

AC: 3x440V/16A; 50/60Hz
DC: 250V/10A

DUT supply
3ph-32A

AC: 3x440V/32A; 50/60Hz
DC: 250V/20A

CRO trigger

5V trigger signal for oscilloscope

MEASUREMENTS

Ring wave

Peak voltage and peak current in LCD

TEST ROUTINES

Quick Start

Immediate start; easy-to-use and
fast

Standard Test
routines

As per IEC 61000-4-12, Level 1 - 4
Manual Standard Test routine
As per ANSI/IEEE C62.41
As per IEC 61000-4-10, Level 1 - 5

User Test routines

Change polarity after n pulses
Change coupling after n pulses
Change voltage after n pulses
Change phase angle after n pulses

INTERFACE

Serial interface

USB

Parallel interface

IEEE 488, addresses 1 - 30

GENERAL DATA

Dimensions,
weight

19"/6HU, approx. 28kg (1-phase)
19"/6HU, approx. 33kg (3-phase)

Supply voltage

115/230V +10/-15%

Fuses

2 x T 2AT (230V); 2 x T4AT (115V)

www.emtest.com