Atec EM-Test-Dito User Manual
Page 2
DATA SHEET > dito > 20100706
TECHNICAL DETAILS
TEST LEVELS
CD mode
500V - 10,000V +/-5%
AD mode
500V - 16,500V +/-5%
Resolution
100V steps
Holding time
More than 5 seconds
Polarity
Positive, negative
Test voltage
measured, indicated in the display
TRIGGER
Single mode
Release of single pulses
Continuous mode
Continuous discharge up to
20Hz (AD) and 10Hz (CD)
Pulse release via
software
Remote control from computer via
fiber optic link (USB)
OUTPUT
CD mode
EN/IEC 61000-4-2; 150pF/330ohm
Rise time
0.8ns +/- 25%
Current 1st peak
7.5A +/- 15% @2kV
15.0A +/- 15% @4kV
22.5A +/- 15% @6kV
30.0A +/- 15% @8kV
Current at 30ns
4.0A +/- 30% @2kV
8.0A +/- 30% @4kV
12.0A +/- 30% @6kV
16.0A +/- 30% @8kV
Current at 60ns
2.0A +/- 30% @2kV
4.0A +/- 30% @4kV
6.0A +/- 30% @6kV
8.0A +/- 30% @8kV
TEST ROUTINES
Easy Zap
Test based on standard levels
Quick Zap
All test parameters are adjustable
during test; fast and easy-to-use
Evaluation
Air discharge with 20Hz; for test
point evaluation on the DUT
Standard Test
routines
As per EN/IEC 61000-4-2,
EN 61000-6-1/-6-2
Set-up
Parameters, standard levels, etc.
DISPLAY AND CONTROLS
Display
LCD graphic display; 128 x 64 pixels
Operation
5 function keys, 2 cursor keys, pulse
trigger
OPERATION
Manual
Via function keys
Remote control
Via built-in optical interface
VOLTAGE SUPPLY
Battery (ACCU)
More than 50,000 pulses @ 16.5kV;
automatic power-off after 10 minutes
if not operated
Mains supply
Via mains supply adapter (MAD)
GENERAL DATA
Weight
940 gramms incl. battery
Case material
ABS; non-inflammable as
per UL 94VO
Temperature
5°C - 40°C (operation conditions)
Rel. humiditiy
20% - 80% (operation conditions)
Battery charger
100V - 240V +/-10%; 50/60Hz
Charging time less than 1.5 hours
ACCESSORIES
DM1
150pF/330ohm discharge module
with built-in contact discharge relay;
Discharge tips for AD and CD
ACCU
Rechargeable battery
ACH
Battery charger
Transportation case, grounding lead
2m, test certificate and operation
manual
www.emtest.com
© EM TEST > PAGE 2/4