3
Second Harmonic
Band 1
Band 2
Band 3*
Intercept Point
9 kHz to
20 MHz to
1 to
8542E/8546A
50 MHz
2.9 GHz
6.5 GHz
100 kHz
≤
f
0
≤
1.8 GHz, > 2.9 GHz
Preamp off
122 dBµV
122 dBµV
134 dBµV
Preamp on
110 dBµV
110 dBµV
100 dBµV
1.8 GHz < f
0
≤
2.9 GHz
Preamp off
105 dBµV
Preamp on
105 dBµV
85422E/85462A
(No bands)
f
0
> 200 kHz
Preamp off
134 dBµV
Preamp on
100 dBµV
Amplitude Specifications (continued)
Other Input Related Spurious
–65 dBc (Band 1, Band 2, and Band 3*)
Residual Responses
(0 dB attenuation, 50
Ω
input termination, preamp on)
8542E/8546A
< 30 kHz
< –2 dBµV
> 30 kHz
< –10 dBµV
85422E/85462A
9 to 150 kHz
< +2 dBµV
150 kHz to 2.9 (or 6.5 GHz*)
< –8 dBµV
Displayed Average Noise Level
(input terminated, 0 dB attenuation, 50
Ω
input termination, 30 Hz IF BW, sample detection
30 Hz averaging BW)
8542E/8546A
30
20
10
0
–10
–20
–30
–40
9
50
100
400
kHz
dB
µ
V
Preamp off
Preamp on
20
10
–20
–30
–40
–50
–10
0
9
100
50
400
kHz
dB
µ
V
Preamp off
Preamp on
f
0
≤
400 kHz
85422E/85462A
f
0
≤
400 kHz
f
0
> 400 kHz
f
0
> 400 kHz
Band 1
Band 2
Band 3*
9 kHz to
20 MHz to
1 to
50 MHz
2.9 GHz
6.5 GHz
≤
–31 dBµV
≤
–31 dBµV
≤
–16 dBµV
≤
–39 dBµV
≤
–39 dBµV
≤
–37 dBµV
Preamp off
Preamp on
Preamp off
≤
–18 dBµV
Preamp on
≤
–39 dBµV
* For 8546A EMI receiver only