beautypg.com

Atec Agilent-11691D User Manual

Page 3

background image

11691D

11692D

Frequency range

2 to 18 GHz

2 to 18 GHz

Minimum directivity

2 to 8 GHz: 30 dB

2 to18 GHz: 30 dB

8 to 18 GHz: 26 dB

1

8 to 18 GHz: 26 dB

1

Maximum primary line SWR

2 to 12.4 GHz: 1.3

2 to 12.4 GHz: 1.3

12.4 to 18 GHz: 1.40

12.4 to 18 GHz: 1.40

Maximum auxiliary port(s) SWR

1.3

1.3

Nominal coupling (dB)

20

20

Maximum coupling variation with frequency (dB)

±1 dB

±l dB

2

Tracking auxiliary arms

*

±0.7 dB

3

*

Maximum primary line residual loss

< 2 dB

< 1.5 dB

Primary line power

50

Ω average

50

Ω average

Handling capability

250

Ω peak

250

Ω peak

Net weight

1.93 kg (4 1b, 4 oz)

2.72 kg (6 1b, 0 oz)

Dimensions

404.8 mm (15-15/16 in) L

404.8 mm (15-15/16 in) L

133.4 mm (5-1/4 in) H

133.4 mm (5-1/4 in) H

42.9 mm (1-11/16 in)W

42.9 mm (1 -11/16 in)W

11691D connector

Primary line

Option

Input/output

Auxiliary arms

Standard

APC-7/APC-7

N(f)

11692D connector options

Primary line

Auxiliary arms

Option

Input/output

Incident/reflection

11692D-005

N(f)/APC-7

N(f)/N(f)

11692D-001

N(f)/N(f)

N(f)/N(f)

11692D-002

N(f)/N(m)

N(f)/N(f)

11692D-003

N(f)/APC-7

APC-7/APC-7

11692D-004

APC-7/APC-7

APC-7/APC-7

Connectors

The APC-7 and type-N connectors are stainless steel
designed for durability. The type-N connectors con-
form to MIL-C-39012 or MIL-C-71.

3

1. 24 dB with type-N connector on the INPUT PORT (11691D) or TEST PORT (11692D).
2. Incident to test port.
3. With the test port shorted and not including source match ripple.
* Typical tracking between 11691D and 11692D is ±0.07 dB.

Figure 2. Directivity specifications.

Figure 3. Typical coupling curve.