Atec Amplifier-Research-FL7040-Kit User Manual
Page 2
Page 2 of 2
SPECIFICATIONS, FL7040
Amplitude Accuracy (field aligned with sensor axes)
Without correction factors applied ............................. 1.0 dB @ 10 MHz
With correction factors applied .................................. Typical expanded measurement uncertainty (95% confidence interval)
0.8 dB, 2 MHz–1 GHz
1.4 dB, 1 GHz–40 GHz
Response Time/ Sampling Rate (through FI7000) ............... 20 msec/up to 50 samples per second, USB and GPIB only
Isotropic Deviation (measured at the critical angle) ............. ±0.5 dB @ 10 MHz
±1.5 dB, 2 MHz–40 GHz (typical)
Operating Range ............................................................ 2–1000 V/m
Linearity, 2 to 1000 V/m .................................................. 0.5 dB
Temperature Stability.. ..................................................... +0.5 dB over operating temperature range
Damage Level ................................................................. 1200 V/m CW
Ranges ........................................................................... Single continuous range
Data returned from probe ................................................ X, Y, Z axes, and composite
Power Requirements ........................................................ Laser powered from FI7000 interface
Dimensions ..................................................................... Approx. 278 x 65 x 65 mm
Probe Head Diameter ...................................................... 65mm
Weight ........................................................................... 150g approx.
Operating Temperature Range ......................................... 10C to 40C (50F to 104F) @ 5% to 95% RH non-condensing
Fiber Optic Connectors ................................................... Two E2000 compact duplex connectors at 1 meter, includes fiber optic verification
loop.
Calibration Data ............................................................. Accredited Calibration Report (A2LA) supplied with probe