5. Test Current:

20

Ω range: 10 mA AC rms

200

Ω range: 1 mA AC rms

2 K

Ω and 20 KΩ range: 100 µA AC rms

6. Maximum Current Spike Resistance:

20

Ω range: 4 KΩ ± 0.5 KΩ

200

Ω range: 25 KΩ ± 3 KΩ

2 K

Ω and 20 KΩ ranges: 50 KΩ ± 5 KΩ

7. Maximum Potential Spike Resistance:

20

Ω range: 10 KΩ ±1 KΩ

200

Ω range: 25 KΩ ± 3 KΩ

2 K

Ω and 20 KΩ ranges: 100 KΩ ±10 KΩ

8. Maximum Output Voltage: 50 V