beautypg.com

Specifications, Panel description, Application ordering information – Atec Chroma-19572 User Manual

Page 2

background image

All specifications are subject to change without notice

Model

19572

Model

Ground Bond

Grounding Resistance Test

Output Current

AC : 3~45A

Load Regulation

1 % + 0.3 A

Resolution

3~30A 0.01A / 30.1~45A 0.1A

Current Accuracy

±

(1.5% of setting + 0.5% of full scale)

Output Frequency

50Hz / 60Hz

Resistance Range

0.1 ~ 510 mΩ

Resistance Resolution

(R display counts/ I display counts)≧0.2,Resolution : 1mΩ
(R display counts/ I display counts) < 0.2, Resolution : 0.1mΩ

Resistance Accuracy

±

(2% of reading + 0.5% of full scale)

Offset

A predetermined value can be subtracted from the measured value and the result of subtraction can

be display

The result of subtraction can be compared with a Good/NO Good judgment reference value,and the

result of comparison can be use for the Good/NO Good judgment

Offset Range

0~100.0mΩ

Test Time

0.5 ~ 999.0 sec., continue

Waveform

Sine wave

GO/NG Judgment

A no-good judgment is made when a resistance greater than the high limit value Is detected.

A no-good judgment is made when the output current is cutout and a no-good Alarm signal is

delivered.

If no abnormal state is detected during the test time, a good judgment is made and a good signal is

deliver.

Limit

HI-LIMIT : 0.1 ~ 510.0mΩ
LOW-LIMIT : OFF, 0.1mΩ ~ HI- LIMIT Value 510.0mΩ MAX.

General

Operation Environment

0 to 40ºC (32 to 104ºF), ≦80% RH.

Power Consumption

No load : < 100VA, With rated load : 880W MAX.

Power Requirement

AC 100V, 120V, 220V-10%, 240V-10% ~ +5%

Weight

< 16 kg.

Dimension(W x H x D)

320 x 105 x 400 mm

SPECIFICATIONS

5 6 8 9

3

2

1

4

14 13 12 11

10
7

16

15

17

18

19

20

21

PANEL DESCRIPTION

14 13 12 11

Verfying integrity of ground connection on appliances and instruments

which use 3 prong power cord

Tests for chassis to ground resistance

Electric of medical applications

Extensive coverage for the testing of UL,CSA,VDE,IEC and other

agency safety requirement

19572 : Ground Bond Tester
A190701 : Remote Control Box
A195720 : GPIB Interface
A190510 : PRINTER Interface

APPLICATION

ORDERING INFORMATION

Distributed by:

Worldwide Distribution and Service Network

19572-201108-500

1. Line Switch
2. Start Key
3. Stop Key
4. Status List
5. Function Keys
6. Cal Enable
7. Update Enable
8. Pass indicator
9. Fail indicator
10. Test indicator

11. Drive (-)
12. Sense (-)
13. Sense (+)
14. Drive (+)
15. Remote I/O
16. Line Voltage Selector
17. Power Cord Receptacle
18. RTN/LOW
19. RS-232 Interface
20. Fan
21. Remote interface

JAPAN

CHROMA JAPAN CORP.

472 Nippa-cho, Kouhoku-ku,

Yokohama-shi, Kanagawa,

223-0057 Japan

http://www.chroma.co.jp
E-mail: [email protected]

U.S.A.

CHROMA SYSTEMS

SOLUTIONS, INC.

25612 Commercentre Drive,

Lake Forest, CA 92630-8830

Tel: +1-949-600-6400

Fax: +1-949-600-6401

Toll Free: +1-866-600-6050

http://www.chromausa.com

E-mail: [email protected]

Developed and Manufactured by :

CHROMA ATE INC.

致茂電子股份有限公司

HEADQUARTERS

No. 66, Hwa-Ya 1st Rd.,

Hwa-Ya Technology Park,

Kuei-Shan Hsiang,33383

Taoyuan County, Taiwan

Tel: +886-3-327-9999

Fax: +886-3-327-8898

http://www.chromaate.com

E-mail: [email protected]

EUROPE

CHROMA ATE EUROPE B.V.

Morsestraat 32, 6716 AH Ede,

The Netherlands

Tel: +31-318-648282

Fax: +31-318-648288

http://www.chromaeu.com

E-mail: [email protected]

CHINA

CHROMA ELECTRONICS

(SHENZHEN) CO., LTD.

8F, No.4, Nanyou Tian An

Industrial Estate, Shenzhen,

China PC: 518052

Tel: +86-755-2664-4598

Fax: +86-755-2641-9620