Specifications, Panel description, Application ordering information – Atec Chroma-19572 User Manual
Page 2
All specifications are subject to change without notice
Model
19572
Model
Ground Bond
Grounding Resistance Test
Output Current
AC : 3~45A
Load Regulation
1 % + 0.3 A
Resolution
3~30A 0.01A / 30.1~45A 0.1A
Current Accuracy
±
(1.5% of setting + 0.5% of full scale)
Output Frequency
50Hz / 60Hz
Resistance Range
0.1 ~ 510 mΩ
Resistance Resolution
(R display counts/ I display counts)≧0.2,Resolution : 1mΩ
(R display counts/ I display counts) < 0.2, Resolution : 0.1mΩ
Resistance Accuracy
±
(2% of reading + 0.5% of full scale)
Offset
■
A predetermined value can be subtracted from the measured value and the result of subtraction can
be display
■
The result of subtraction can be compared with a Good/NO Good judgment reference value,and the
result of comparison can be use for the Good/NO Good judgment
Offset Range
0~100.0mΩ
Test Time
0.5 ~ 999.0 sec., continue
Waveform
Sine wave
GO/NG Judgment
■
A no-good judgment is made when a resistance greater than the high limit value Is detected.
■
A no-good judgment is made when the output current is cutout and a no-good Alarm signal is
delivered.
■
If no abnormal state is detected during the test time, a good judgment is made and a good signal is
deliver.
Limit
HI-LIMIT : 0.1 ~ 510.0mΩ
LOW-LIMIT : OFF, 0.1mΩ ~ HI- LIMIT Value 510.0mΩ MAX.
General
Operation Environment
0 to 40ºC (32 to 104ºF), ≦80% RH.
Power Consumption
No load : < 100VA, With rated load : 880W MAX.
Power Requirement
AC 100V, 120V, 220V-10%, 240V-10% ~ +5%
Weight
< 16 kg.
Dimension(W x H x D)
320 x 105 x 400 mm
SPECIFICATIONS
5 6 8 9
3
2
1
4
14 13 12 11
10
7
16
15
17
18
19
20
21
PANEL DESCRIPTION
14 13 12 11
■
Verfying integrity of ground connection on appliances and instruments
which use 3 prong power cord
■
Tests for chassis to ground resistance
■
Electric of medical applications
■
Extensive coverage for the testing of UL,CSA,VDE,IEC and other
agency safety requirement
19572 : Ground Bond Tester
A190701 : Remote Control Box
A195720 : GPIB Interface
A190510 : PRINTER Interface
APPLICATION
ORDERING INFORMATION
Distributed by:
Worldwide Distribution and Service Network
19572-201108-500
1. Line Switch
2. Start Key
3. Stop Key
4. Status List
5. Function Keys
6. Cal Enable
7. Update Enable
8. Pass indicator
9. Fail indicator
10. Test indicator
11. Drive (-)
12. Sense (-)
13. Sense (+)
14. Drive (+)
15. Remote I/O
16. Line Voltage Selector
17. Power Cord Receptacle
18. RTN/LOW
19. RS-232 Interface
20. Fan
21. Remote interface
JAPAN
CHROMA JAPAN CORP.
472 Nippa-cho, Kouhoku-ku,
Yokohama-shi, Kanagawa,
223-0057 Japan
http://www.chroma.co.jp
E-mail: [email protected]
U.S.A.
CHROMA SYSTEMS
SOLUTIONS, INC.
25612 Commercentre Drive,
Lake Forest, CA 92630-8830
Tel: +1-949-600-6400
Fax: +1-949-600-6401
Toll Free: +1-866-600-6050
http://www.chromausa.com
E-mail: [email protected]
Developed and Manufactured by :
CHROMA ATE INC.
致茂電子股份有限公司
HEADQUARTERS
No. 66, Hwa-Ya 1st Rd.,
Hwa-Ya Technology Park,
Kuei-Shan Hsiang,33383
Taoyuan County, Taiwan
Tel: +886-3-327-9999
Fax: +886-3-327-8898
http://www.chromaate.com
E-mail: [email protected]
EUROPE
CHROMA ATE EUROPE B.V.
Morsestraat 32, 6716 AH Ede,
The Netherlands
Tel: +31-318-648282
Fax: +31-318-648288
http://www.chromaeu.com
E-mail: [email protected]
CHINA
CHROMA ELECTRONICS
(SHENZHEN) CO., LTD.
8F, No.4, Nanyou Tian An
Industrial Estate, Shenzhen,
China PC: 518052
Tel: +86-755-2664-4598
Fax: +86-755-2641-9620