beautypg.com

Specifications – Atec AEMC-4620_4630 User Manual

Page 3

background image

Models 4620 & 4630 (pdf)

08/02

Technical Assistance (800) 343-1391

www.aemc.com

3 of 6

Specifications

MODELS

4620

4630

ELECTRICAL

Ranges (Auto-Ranging 0 to 2000Ω)

20Ω

200Ω

2000Ω

Measurement

0.00 to 19.99Ω

20.0 to 199.9Ω

200 to 1999Ω

Resolution

10mΩ

100mΩ

1Ω

Open Voltage

≤42V peak

≤42V peak

≤42V peak

Resistance Measurement

128Hz square wave

128Hz square wave

128Hz square wave

Frequency
Test Current

10mA

1mA

0.1mA

Accuracy

±2% of Reading ± 1ct

±2% of Reading ± 1ct

±2% of Reading ± 3cts

Auxiliary Electrode Influence

Max Res. in Current Circuit

3kΩ

30kΩ

50kΩ

Max Res. in Voltage Circuit

2000Ω

50kΩ

50kΩ

Interference

Rejects high levels of interference voltage (DC, 50 to 60Hz, harmonics):

DC voltage in series with X: 20V; AC voltage in series with Y: 13V peak;

AC voltage in series with Z: 32V peak

Response Time

Approximately 4 to 8 seconds for a stabilized measurement

Withstanding Voltage

250V

AC

or 100V

DC

Power Source

Eight C cell batteries;

120/230V 50/600Hz

Alkaline recommended;

Rechargeable 9.6V, 3.5 Ah NiMH batteries

Battery Life

1800 15-second measurements; LO BAT indication on LCD

Fuse Protection

0.1A, >250V, 0.25 x 1.25"; 30kA Interrupt Capacity

MECHANICAL

Display

3

1

/

2

digit, 0.71" (18mm) high; 2000-counts; electroluminescent blue backlight

LCD also indicates overrange, test lead shorts and lead reversals

Connection

Color-coded terminals accept spade lugs with min. gap of 6mm or

standard 4mm banana jacks

LED Indication

Three LEDs indicate high input noise, high auxiliary rod resistance,

open leads, blown fuse

Operating Temperature

14° to 131°F (-10° to 55°C), 20 to 90% RH

Storage Temperature

-40° to 158°F (-40° to 70°C), 10 to 90% RH with batteries removed

Dimensions

10.8 x 9.7 x 5.0" (273 x 247 x 127mm)

Weight

6.28 lbs (2.85kg)

7.38 lbs (3.35kg)

Case

Heavy duty o-ring sealed field case

Mechanical Shock

IEC 68-2-27

Vibration Test

IEC 68-2-6

6

Drop Test

IEC 68-2-32

Dielectric Test

3kV, 50/60Hz, 1 min. between four interconnected measuring terminals

and any external metal ground

Environmental

O-ring sealed against dust and water to IP50 (Protection Index)

when case is closed

SAFETY

Rating

EN 61010-1, Cat. III, Pollution Degree 2

CE Mark

Yes