beautypg.com

Specifications, Model 5070 – Atec AEMC-5070 User Manual

Page 5

background image

Specifications

MoDeL

5070

ELECTRICAL

Insulation Tests

Test Voltage/Range

500V

30k

Ω to 2000GΩ (2TΩ)

1000V

100k

Ω to 4000GΩ (4TΩ)

2500V

100k

Ω to 10,000GΩ (10TΩ)

5000V

300k

Ω to 10,000GΩ (10TΩ)

User Selectable Test Voltage

Programmable:

40 to 1000V: 10V increments;

1000 to 5100V: 100V increments

Three test voltages can be stored.

Short Circuit Current

<1.6mA ± 5%

Accuracy

1k

Ω to 40GΩ

±5% of Reading ± 3cts

40G

Ω to 10TΩ

±15% of Reading ± 10cts

DAR (1 min/30 sec)

(1)

0.02 to 50.00

PI (10 min/1 min & user programmable)

0.02 to 50.00

DD

Current after 1 min

)

0.02 to 50.00

test voltage x capacitance

Capacitance Check

0.005 to 49.99µF

Max resolution 1nF

Leakage Current Measurement

0.00nA to 3mA

Max resolution 1pA

Programmable Run Time R(t)

1 min to 49 hrs 59 min

Smooth Function (user selectable)

Digital filtering stabilizesdisplay readings

Automatic Step Voltage

3 profiles, each containing 5 steps programmable from 40 to 5100V.

Dwell time programmable from 1 min to 10 hrs per step.

Maximum total time for 5 steps is 59 hrs 59 min

Temperature Correction

Operator programmable reference temperature and device under

test temperature with automatic temperature correction of resistance

Discharge After Test

Automatic

Discharge Voltage Display

Yes

Voltage Test/Safety Check

0 to 1000V

ac

/

dc

(16 to 420H

z

), 1V Resolution

Voltage Warning Indicator

Yes >25V

Test Inhibition

(2)

Yes >25V

Guard Terminal

Yes – adjustable from 25 to 1000V

depending on test voltage range in use

Power Source

Eight NiMH rechargeable batteries

Line power: 85 to 256V

ac

(50/60Hz)

MEChAnICAL

Dimensions

10.63 x 9.84 x 7.09" (270 x 250 x 180mm)

Weight

9.5 lbs (4.3kg)

Protection Index

IP53

DISPLAy

Backlight

Blue electroluminescent

Display Size

3.75 x 3" (93 x 75mm)

Display

Graphical: 320 x 240 resolution

CoMMunICATIon

Report Print out Direct to Printer

Yes – preset format

Storage of Readings over Time R(t)

128kB memory

Programmable Reading Intervals

5 sec to 15 min

Test Voltage Display

Yes

Elapsed Test Time Display

Yes

Real Time/Date Display

Yes

Test Voltage Lock-out

User programmed

Storage of Test Results

Stores over 1500 test results

Communication Port

RS-232

PC Software/Report Generation

Yes, DataView

®

(included)

PC operation of Megohmmeter

Yes

SAfETy

Safety Rating

EN 61010-1, 1000V Cat. III

Double Insulation

Yes

CE Mark

Yes

(1)

DAR times are programmable in the Model 5070

(2)

Inhibit voltage is selectable at 3, 10 or 20% of test voltage

(

)

5000V Graphical Megohmmeter Model 5070 (pdf) Rev. 06 02/08

Technical Assistance (800) 343-1391

www.aemc.com 5 of 11