The D30R is a stand alone unit specifically designed to

microseconds), so the D30R complies with IEEE Standard

diagnose faults in very large electrical motors and windings,

522-1992 and IEC Standard 34-15 when testing motor

improving quality in the shop and reducing unnecessary

windings and coils.

and costly downtime in the field. The 30KV output allows

The D30R is housed in a new mobile case with the control

you to thoroughly test larger windings with lower

unit permanently affixed to the upper face. 60Kv high

impedance and higher capacitance. The D30R satisfies

voltage test leads are provided along with dedicated

the requirements of testing the windings of both AC motors

Kelvin Resistance test leads for convenient portable testing.

and DC armatures by producing a Surge with higher

These features along with the unsurpassed testing

voltage and instantaneous current.

capabilities make the D30R a powerful and technically

The D30R offers you all of the convenient features of digital

advanced tester for in house shop or field environments.

technology. It performs Resistance, DC High Potential and

Surge tests along with incorporating a supply monitor to

insure safe operation from a well grounded source. The

control and display module provides the user with

comprehensive testing results.

The Resistance Test verifies the existence of dead shorts

within the turn-to-turn coils, shows any imbalances between

phases due to turn count differences, along with locating

poor wire connections or contacts.
The DC high potential (HiPot) test is also done with the

D30R. Test voltage is set by the output control from 500

volts up to 30,000 volts. Current is displayed and an

overcurrent trip circuit monitors the test. If current exceeds

the trip level, the test is automatically halted. In its most

sensitive setting, the protective circuit will operate as low as

10 microamps.

The “Key Element” of the Surge test circuit is a grounded

grid hydrogen thyratron tube. The tube conducts very high

peak current (rated up to 20,000 amps). Such high

instantaneous currents are needed to fully test the very low

impedance windings of large motors or form wound coils.

This current is only produced for a few microseconds by the

discharge of a 0.1 microfarad capacitor.

Voltage rise time is 100-200 nanoseconds (0.1 - 0.2

BAKER INSTRUMENT COMPANY

PO Box 587, Fort Collins, CO 80522
PH: (970) 282-1200, (800) 752-8272
Fax: (970) 282-1010
Url: www.bakerinst.com
·Fort Collins ·Nurnberg

SPECIFICATIONS*

SURGE TEST
Maximum Output Voltage

30,000 Volts

Maximum Output Current

1,800-2,000 amps peak

Maximum Pulse Energy

45 joules

Discharge Capacitance

.1 micro-farad

DC HIGH POTENTIAL TEST
Maximum Output Voltage

30,000 Volts

Maximum Output Current

1000 microamps

Overcurrent Trip

10/100/1000 microamps

Current Resolution

1/10/100 microamps

RESISTANCE TEST

.0008 ohms - 216 ohms

PHYSICAL CHARACTERISTICS
Weight (pounds)

290 pounds

Dimensions

24 x 55 x 26 inches

Power Requirements

110V/220V Single Phase
1000 Watts, 50/60 Hz

Three Phase Selector Switch Standard on this unit.

* Data subject to change without notice. Printed in USA 10/03