beautypg.com

Atec AEMC-8335 User Manual

Page 7

background image

6

PowerPad Model 8335

Accuracy Specifications of the Current Probes (to be added to the meter’s accuracy)

Sensor Type

TRMS Current

Maximum Error on I

RMS

Maximum Error on

MR193

1000A

[1A; 10A]

±(1.5%+1A)

N.S.

[10A; 100A]

±(2°)

[100A; 800A]

±(3%)

±(1.5°)

[800A; 1200A]

±(5%)

[1200A; 1400A]

SR193

1000A

[1A; 3A]

±(0.8%)

N.S.

[3A; 10A]

±(1°)

[10A; 100A]

±(0.3%)

±(0.5°)

[100A; 1200A]

±(0.2%)

±(0.3°)

AmpFlex

®

3000A

[10A; 100A]

±(3%)

±(1°)

[100A; 6500A]

±(2%)

±(0.5°)

MiniFlex

1000A

[10A; 100A]

±(3%)

±(1°)

[100A; 1000A]

±(2%)

±(0.5°)

MN93

200A

[0,5A; 2A]

±(3%+1A)

N.S.

[2A; 10A]

±(6°)

[10A; 100A]

±(2.5%+1A)

±(3°)

[100A; 240A]

±(1%+1A)

±(2°)

MN193

100A

[100mA; 300mA]

±(0.7%+2mA)

N.S.

[300mA; 1A]

±(1.5°)

[1A; 120A]

±(0.7%)

±(0.7°

MN193

5A

[5mA; 50mA]

±(1%+0.1mA)

±(1.7°)

[50mA; 500mA]

±(1%)

±(1°)

[500mA; 6A]

±(0.7%)

5A

Adapter

[5mA; 50mA]

±(1%)

±(1°)

[50mA; 6A]

±(0.5%)

±(0°)