Input Specifications
Voltage
115/230 VAC ± 10%, user selectable
Frequency
50/60 Hz ± 5%
Fuse
3.15 A, fast acting 250 VAC
Dielectric Withstand Test Mode
Output Rating
5000 V @ 20 mA AC
6000 V @ 7.5 mA DC
Voltage Setting
Range: 0 - 5.00 KV AC
0 - 6.00 KV DC
Resolution: 0.01 KV
Accuracy: ± (2% of setting + 5 V)
Maximum Limit
AC
Range: 0.00 - 20.00 mA
Resolution: 0.01 mA
DC
Range: 0 - 7500 µA
Resolution: 1 µA
Accuracy: AC and DC ± (2% of setting + 2 counts)
Minimum Limit
AC
Range: 0.000 - 9.999 mA
Resolution: 0.001 mA
DC
Range: 0.0 - 999.9 µA
Resolution: 0.1 µA
Accuracy: AC and DC ± (2% of setting + 2 counts)
Arc Detection
Range: 0 - 9, 0 disabled
Ground Fault Interrupt
GFI Trip Current: 450 µA max (AC or DC)
HV Shut Down Speed: < 1ms
Current Display
Auto Range
AC
Range 1: 0.000 - 3.500 mA
Range 2: 3.00 - 20.00 mA
DC
Range 1: 0.0 µA - 350.0 µA
Range 2: 0.300 mA - 3.500 mA
Range 2: 3.00 mA - 7.50 mA
Accuracy
All Ranges ± (2% of reading + 2 counts)
DC Output Ripple
≤ 5% Ripple RMS at 6 KV DC @ 7.5 mA, Resistive Load
Discharge Time
≤ 200 ms
The maximum capacitive load vs output voltage:
0.20 µF < 1 KV
0.050 µF < 4 KV
0.10 µF < 2 KV
0.040 µF < 5 KV
0.06 µF < 3 KV
0.015 µF < 6 KV
AC Voltage Waveform
Sine Wave, Crest Factor = 1.3 - 1.5, <2% THD
Output Frequency
Range: 50 or 60 HZ, User Selectable
Output Voltage Regulation ± (1% of output + 5 V) from no load to full load and over
input voltage range.
Dwell Timer
Range: AC 0, 0.3 - 999.9 sec (0 = Constant)
DC 0, 0.4 - 999.9 sec (0 = Constant)
Accuracy: ± (0.1% of reading + 0.05 sec)
Ramp Timer
Range: Ramp-Up: 0.1 - 999.9 sec
Ramp-Down: AC 0.0 - 999.9 sec
DC 1.0 - 999.9 sec (0=OFF)
Accuracy: ± (0.1% of reading + 0.05 sec)
Ground Continuity Current
DC 0.1 A ± 0.01 A, fixed
Ground Continuity
Range: 0.0
Ω - 1.50 Ω
Maximum Limit
Resolution: 0.01
Ω
Minimum Limit
Accuracy: ± (3% of setting + 0.02
Ω)
Ground Continuity
Range: 0.0
Ω - 0.50 Ω
Auto Offset
Resolution: 0.01
Ω
Accuracy: ± (3% of setting + 0.02
Ω)
Insulation Resistance Test Mode
Voltage Setting
Range: 30 - 1000 VDC
Accuracy: ± (2% of setting + 5 V)
Resistance Display
Range: 1 - 9999 M
Ω (4 Digit, Auto Ranging)
Resolution: 500 VDC - 1000 VDC
M
Ω
M
Ω
0.001
1.000 - 9.999
0.01
10.00 - 99.99
0.1
100.0 - 999.9
1
1000 - 9999
Accuracy: ± (2% of reading +2 counts) at test voltage
500 - 1000 V and 1 - 999.9 M
Ω
± (5% of reading +2 counts) at test voltage
500 - 1000 V and 1000 - 9999 M
Ω
± (8% of reading +2 counts) at test voltage
30 - 500 V and 1 - 1000 M
Ω
Maximum Limit
Range: 0, 1 - 9999 M
Ω (0=OFF)
Accuracy: Same as Resistance Display
Minimum Limit
Range: 1 - 9999 M
Ω
Accuracy: Same as Resistance Display
Ramp Timer
Range: Ramp-Up: 0.1 - 999.9 sec
Ramp-Down: 1.0 - 999.9 sec (0=OFF)
Accuracy: ± (0.1% of reading + 0.05 sec)
Delay Timer
Range: 0, 0.5 - 999.9 sec (0 = Constant)
Accuracy: ± (0.1% of reading + 0.05 sec)
Ground Fault Interrupt
GFI Trip Current: 450 µA max
HV Shut Down Speed: < 1 ms
General Specifications
Remote Control &
The following input and output signals are provided
Signal Output
through two 9 pin D type connectors:
1. Remote control: Test, Reset, and Remote Interlock
2. Remote recall of memor y program #1, #2, and #3
3. Outputs: Pass, Fail, Test-in-process, and Reset
Program Memory
10 Memories, 3 steps per memor y.
Security
Key Lock capability to avoid unauthorized access to
all test parameters and memor y locations.
Mechanical
Bench or rack mount with tilt up feet.
Dimensions
(W x H x D) 8.46 x 3.5 x 14.57 in.
(215 x 89 x 370 mm)
Weight
20.96 lbs (9.53 kgs)
Calibration
Traceable to National Institute of Standards &
Technology (NIST). Calibration controlled by software.
Adjustments are made through front panel keypad in
a restricted access calibration mode. Calibration
information stored in non-volatile memor y.
Available Accessories
Probe (38081)
High voltage retractable 5 KV AC probe.
Return Probe (38082)
Retractable probe used on the return side of the Hypot III.
Safe-T-Probe (38083)
Test Gun with trigger that controls retractable probe and
activates the instrument’s high voltage circuit.
Footswitch (35822)
Interlocked footswitch that provides a “hands-off”
remote start.
At Associated Research, Safety Compliance Testing Is Our Only Focus.
13860 W. Laurel Dr., Lake Forest, IL USA 60045-4546
■
P: +1.847.367.4077
■
F: +1.847.367.4080
■
www.asresearch.com
1.800.858.TEST [8378]
■
Customer support & technical ser vices
■
5-Year extended warranty*
■
24-Hour turn-around on calibrations
■
Industr y seminars, expert training & education programs
■
Local sales offices throughout the world
*With annual calibration
from Associated Research.
Specifications subject to change without notice.
Recycled Equipment (410)685-1997 www.recycledequipment.com email: [email protected]