beautypg.com

Input specifications, Dielectric withstand test mode, Insulation resistance test mode – Atec Associated-Research-3605-65-70 User Manual

Page 2: General specifications, Available accessories

background image

Input Specifications

Voltage

115/230 VAC ± 10%, user selectable

Frequency

50/60 Hz ± 5%

Fuse

3.15 A, fast acting 250 VAC

Dielectric Withstand Test Mode

Output Rating

5000 V @ 20 mA AC
6000 V @ 7.5 mA DC

Voltage Setting

Range: 0 - 5.00 KV AC

0 - 6.00 KV DC

Resolution: 0.01 KV
Accuracy: ± (2% of setting + 5 V)

Maximum Limit

AC

Range: 0.00 - 20.00 mA
Resolution: 0.01 mA

DC

Range: 0 - 7500 µA
Resolution: 1 µA
Accuracy: AC and DC ± (2% of setting + 2 counts)

Minimum Limit

AC

Range: 0.000 - 9.999 mA
Resolution: 0.001 mA

DC

Range: 0.0 - 999.9 µA
Resolution: 0.1 µA
Accuracy: AC and DC ± (2% of setting + 2 counts)

Arc Detection

Range: 0 - 9, 0 disabled

Ground Fault Interrupt

GFI Trip Current: 450 µA max (AC or DC)
HV Shut Down Speed: < 1ms

Current Display

Auto Range

AC

Range 1: 0.000 - 3.500 mA
Range 2: 3.00 - 20.00 mA

DC

Range 1: 0.0 µA - 350.0 µA
Range 2: 0.300 mA - 3.500 mA
Range 2: 3.00 mA - 7.50 mA

Accuracy

All Ranges ± (2% of reading + 2 counts)

DC Output Ripple

≤ 5% Ripple RMS at 6 KV DC @ 7.5 mA, Resistive Load

Discharge Time

≤ 200 ms
The maximum capacitive load vs output voltage:

0.20 µF < 1 KV

0.050 µF < 4 KV

0.10 µF < 2 KV

0.040 µF < 5 KV

0.06 µF < 3 KV

0.015 µF < 6 KV

AC Voltage Waveform

Sine Wave, Crest Factor = 1.3 - 1.5, <2% THD

Output Frequency

Range: 50 or 60 HZ, User Selectable

Output Voltage Regulation ± (1% of output + 5 V) from no load to full load and over

input voltage range.

Dwell Timer

Range: AC 0, 0.3 - 999.9 sec (0 = Constant)

DC 0, 0.4 - 999.9 sec (0 = Constant)

Accuracy: ± (0.1% of reading + 0.05 sec)

Ramp Timer

Range: Ramp-Up: 0.1 - 999.9 sec

Ramp-Down: AC 0.0 - 999.9 sec

DC 1.0 - 999.9 sec (0=OFF)

Accuracy: ± (0.1% of reading + 0.05 sec)

Ground Continuity Current

DC 0.1 A ± 0.01 A, fixed

Ground Continuity

Range: 0.0

Ω - 1.50 Ω

Maximum Limit

Resolution: 0.01

Minimum Limit

Accuracy: ± (3% of setting + 0.02

Ω)

Ground Continuity

Range: 0.0

Ω - 0.50 Ω

Auto Offset

Resolution: 0.01

Accuracy: ± (3% of setting + 0.02

Ω)

Insulation Resistance Test Mode

Voltage Setting

Range: 30 - 1000 VDC
Accuracy: ± (2% of setting + 5 V)

Resistance Display

Range: 1 - 9999 M

Ω (4 Digit, Auto Ranging)

Resolution: 500 VDC - 1000 VDC

M

M

0.001

1.000 - 9.999

0.01

10.00 - 99.99

0.1

100.0 - 999.9

1

1000 - 9999

Accuracy: ± (2% of reading +2 counts) at test voltage

500 - 1000 V and 1 - 999.9 M

± (5% of reading +2 counts) at test voltage
500 - 1000 V and 1000 - 9999 M

± (8% of reading +2 counts) at test voltage
30 - 500 V and 1 - 1000 M

Maximum Limit

Range: 0, 1 - 9999 M

Ω (0=OFF)

Accuracy: Same as Resistance Display

Minimum Limit

Range: 1 - 9999 M

Accuracy: Same as Resistance Display

Ramp Timer

Range: Ramp-Up: 0.1 - 999.9 sec

Ramp-Down: 1.0 - 999.9 sec (0=OFF)

Accuracy: ± (0.1% of reading + 0.05 sec)

Delay Timer

Range: 0, 0.5 - 999.9 sec (0 = Constant)
Accuracy: ± (0.1% of reading + 0.05 sec)

Ground Fault Interrupt

GFI Trip Current: 450 µA max
HV Shut Down Speed: < 1 ms

General Specifications

Remote Control &

The following input and output signals are provided

Signal Output

through two 9 pin D type connectors:
1. Remote control: Test, Reset, and Remote Interlock
2. Remote recall of memor y program #1, #2, and #3
3. Outputs: Pass, Fail, Test-in-process, and Reset

Program Memory

10 Memories, 3 steps per memor y.

Security

Key Lock capability to avoid unauthorized access to
all test parameters and memor y locations.

Mechanical

Bench or rack mount with tilt up feet.

Dimensions

(W x H x D) 8.46 x 3.5 x 14.57 in.

(215 x 89 x 370 mm)

Weight

20.96 lbs (9.53 kgs)

Calibration

Traceable to National Institute of Standards &
Technology (NIST). Calibration controlled by software.
Adjustments are made through front panel keypad in
a restricted access calibration mode. Calibration
information stored in non-volatile memor y.

Available Accessories

Probe (38081)

High voltage retractable 5 KV AC probe.

Return Probe (38082)

Retractable probe used on the return side of the Hypot III.

Safe-T-Probe (38083)

Test Gun with trigger that controls retractable probe and
activates the instrument’s high voltage circuit.

Footswitch (35822)

Interlocked footswitch that provides a “hands-off”
remote start.

At Associated Research, Safety Compliance Testing Is Our Only Focus.

13860 W. Laurel Dr., Lake Forest, IL USA 60045-4546

P: +1.847.367.4077

F: +1.847.367.4080

[email protected]

www.asresearch.com

1.800.858.TEST [8378]

Customer support & technical ser vices

5-Year extended warranty*

24-Hour turn-around on calibrations

Industr y seminars, expert training & education programs

Local sales offices throughout the world

*With annual calibration

from Associated Research.

Specifications subject to change without notice.

Recycled Equipment (410)685-1997 www.recycledequipment.com email: [email protected]