Panel description, Specifications ordering information – Atec Chroma-19056_19057 Series User Manual
Page 2
PANEL DESCRIPTION
All specifications are subject to change without notice
SPECIFICATIONS
ORDERING INFORMATION
1.
Power Switch
2.
Start Key
3.
Stop Key
4.
Function Key
5.
Menu Key
6.
Main Index Key
7.
Local Key
8.
LED Display
9.
HV Output
10.
RTN/LOW
11.
Entry Keys
12.
Interlock
13.
OPTION
14.
SCAN
15.
RS232 Interface
16.
ARC Signal Output
17.
HANDLER Interface
18.
USB Interface
19.
Power Inlet
20.
HV
21.
HV CONTACT
22.
LOW/CONTACT
23.
RTN/LOW
1
2
3
10
9
4
5 6 7
8
11
12 13
14 15
17
18
19
23 22
21 20
19056 : Hipot Analyzer AC10kV
19057 : Hipot Analyzer DC12kV/IR
19057-20 : Hipot Analyzer DC20kV/IR
A190508 : GPIB Interface
A190702 : 40kV High Voltage Testing Rod
A190708 : ARC (Flashover) Verification Fixture
16
Model
19056
19057
19057-20
Mode
ACV
DCV / IR
DCV / IR
Withstanding Voltage Test
Output Voltage
AC: 0.1~10kV
DC: 0.1~12kV
DC : 0.1 ~ 20kV
Load Regulation
±
(1% of output + 10V), Rated load
Voltage Accuracy
1% of reading + 0.1% of full scale
1.5% of reading + 0.1% of full scale
Voltage Regulation
±
(1% of output + 10V), Rated load
Cutoff Current
0.001~20mA
0.0001~10mA
0.0001~5 mA
Current Accuracy
0.100mA~2.999mA :
±
(1% of reading + 0.3% of full range)
3.00mA~20.00mA :
±
(1.5% of reading + 0.3% of full range)
±
(1% of reading + 0.5% of full range)
Current Resolution
AC : 1μA
DC : 0.1 μA
Output Frequency
50Hz / 60Hz
-
-
Test/Ramp/Fall/Dwell Time
0.3 ~ 999 sec., continue / 0.1 ~ 999 sec., off / 0.1 ~ 999 sec., off / 0.1 ~ 999 sec., off
Waveform
Sine wave
-
-
Insulation Resistance Test
Output Voltage
-
DC : 0.1 ~ 5kV
Voltage Resolution
-
2V
Voltage Accuracy
-
1% of setting + 0.5% of full scale
1.5% of setting + 0.5% of full scale
IR Range
-
0.1MΩ ~ 50GΩ
Resistance Resolution
-
0.1MΩ
Resistance
Accuracy
≧
0.5kV
-
1MΩ ~ 1GΩ : ± 3% of reading + 0.5% of full range
1GΩ ~ 10GΩ : ± 5% of reading + 1% of full range
10GΩ ~ 50GΩ : ± 10% of reading + 1% of full range
<0.5kV
-
1MΩ ~ 1GΩ : ± 5% of reading + (0.5*500/Vs)% of full scale
Flashover Detection
Setting Mode
Programmable setting
Detection Current
AC : 20mA
DC : 10mA
DC : 10mA
Contact Check Function
Contact Check
OSC (open/short check)
HVCC(High Voltage contact check)
HFCC (High Frequency Contact Check)
HVCC(High Voltage contact check)
HFCC (High Frequency Contact Check)
HVCC(High Voltage contact check)
HFCC (High Frequency Contact Check)
Electrical Hazard Protection Function
Ground Fault Interrupt
0.5mA±0.25mA AC, ON/OFF
-
-
Key Lock
Yes (password control)
Interlock
YES
GO/NG Judgment Window
Indication, Alarm
GO : Short sound, Green LED; NG : Long sound, Red LED
Memory Storage
100 sets ,max. 50 steps per set
Interface
Standard-RS232, Handler interface ,USB , SCAN Optional - GPIB interface
General
Operation Environment
Temperature: 0℃ ~ 45℃ ; Humidity: 15% to 95% R.H@ ≦ 40℃
Power Consumption
600VA
Power Requirements
100~240Vac, 47~66Hz
Dimension (HxWxD)
130x430x500 mm/5.12x16.93x19.69 inch
Weight
28kg / 61.7 lbs
19056-19057-E-201207-2000
Worldwide Distribution and
Service Network
JAPAN
CHROMA JAPAN CORP.
472 Nippa-cho, Kouhoku-ku,
Yokohama-shi, Kanagawa,
223-0057 Japan
http://www.chroma.co.jp
E-mail: [email protected]
U.S.A.
CHROMA SYSTEMS
SOLUTIONS, INC.
25612 Commercentre Drive,
Lake Forest, CA 92630-8830
Tel: +1-949-600-6400
Fax: +1-949-600-6401
http://www.chromausa.com
E-mail: [email protected]
Developed and Manufactured by :
CHROMA ATE INC.
致茂電子股份有限公司
HEADQUARTERS
66 Hwaya 1st Rd., Kueishan
Hwaya Technology Park,
Taoyuan County 33383,
Taiwan
Tel: +886-3-327-9999
Fax: +886-3-327-8898
http://www.chromaate.com
E-mail: [email protected]
EUROPE
CHROMA ATE EUROPE B.V.
Morsestraat 32, 6716 AH Ede,
The Netherlands
Tel: +31-318-648282
Fax: +31-318-648288
http://www.chromaeu.com
E-mail: [email protected]
CHINA
CHROMA ELECTRONICS
(SHENZHEN) CO., LTD.
8F, No.4, Nanyou Tian An
Industrial Estate, Shenzhen,
China PC: 518052
Tel: +86-755-2664-4598
Fax: +86-755-2641-9620