Specifications – Atec AEMC-6240-6250 User Manual
Page 4
![background image](/manuals/318068/4/background.png)
4
www.aemc.com
Technical Assistance (800) 343-1391
MODELS
6240
6250
ELECTRICAL
Range
4000μ
Ω
40m
Ω
400m
Ω
4000m
Ω
40
Ω
400
Ω
5.000m
Ω 25.000mΩ 250.00mΩ 2500.0mΩ 25.000Ω 250.00Ω 2500.0Ω
Accuracy
0.25% of
Reading
± 2cts
0.25% of
Reading
± 2cts
0.25% of
Reading
± 2cts
0.25% of
Reading
± 2cts
0.25% of
Reading
± 2cts
0.25% of
Reading
± 2cts
0.15% of
Reading +
1μ
Ω
0.05% of
Reading +
3μ
Ω
0.05% of
Reading +
30μ
Ω
0.05% of
Reading +
0.3m
Ω
0.05% of
Reading +
3m
Ω
0.05% of
Reading +
30m
Ω
0.05% of
Reading +
300m
Ω
Resolution
1μ
Ω
10μ
Ω
0.1m
Ω
1m
Ω
10m
Ω
100m
Ω
0.1μ
Ω
1μ
Ω
10μ
Ω
0.1m
Ω
1m
Ω
10m
Ω
100m
Ω
Test Current
10A
1A
1A
100mA
10mA
10mA
10A
10A
10A
1A
100mA
10mA
1mA
Measurement
Mode
Single test start/stop or auto test and store of multiple measurements
Selectable: Inductive (continuous test), Resistive (instantaneous test)
or Auto (multiple tests)
Power Source
Rechargeable 6V, 8.5Ah NiMH batteries
Battery Life
Up to 850, 10A tests
Approximately 5000, 10A tests
Battery
Charging
Internal 110/230V
ac
, 50/60Hz charger
Internal 120/240V
ac
, 45/400Hz line voltage
Power
Management
User selectable power save mode
Display powers down after 1, 5 or 10 minutes of inactivity (operator selectable)
Low Battery
Indication
blinks when battery needs changing
displays when battery needs changing
Memory
Stores up to 99 test results
Data in memory can be reviewed on the instrument display,
on a PC or via direct printout
Stores up to 1500 test results
Data in memory can be reviewed on the instrument display,
on a PC or via direct printout
MECHANICAL
Dimensions
10.70 x 9.76 x 7.17" (270 x 250 x 180mm)
10.63 x 9.84 x 7.09" (270 x 250 x 180mm)
Weight
10 lbs. (4.5kg)
8.1 lbs (3.69kg)
Index of
Protection
IP53 (case open); IP54 (case closed)
IP53 (case open); IP64 (case closed)
DISPLAy
Display Type
4000-count LCD
Large 50,000-count digital
Size
4 x 2.25" (102 x 57mm)
Blacklight
Blue electroluminescent
Color
Safety yellow case; gray faceplate UL94 V-O
COMMuNICATION
Interface Port
Optically isolated USB
a-pin, RS-232
ENvIRONMENTAL
Operating
Temperature
14° to 132°F (-10° to 55°C); 10 to 85% RH
14° to 132°F (-10° to 55°C); 10 to 80% RH (non-condensing)
Storage
Temperature
-40° to +158°F (-40° to +70°C); 10 to 90% RH
-40° to 140°F (-40° to 60°C)
SAfETy
Safety Rating
EN61010-1 300V CAT II
EN 61010-1, 50V CAT III
Double
Insulation
Yes
CE Mark
Yes
Catalog No.
2129.80
2129.81
SPECIFICATIONS