FEATURES
•
8 universal analog input channels
•
8 digital I/Os (individually configured as
inputs or outputs)
•
2 relay outputs (NO, NC and common)
•
RS485 interface
(Modbus master or slave)
•
24V
dc
output to power up to eight
4 to 20mA transmitters
•
Ethernet interface for LAN and
internet use
•
USB-device interface for configuring,
monitoring and download
•
USB-host interface for logged data
retrieval through a USB flash drive
•
Up to 32 configurable alarms
•
Up to 128 virtual channels
•
Detachable display (optional)
•
Up to 100 channels can be logged at a
configurable rate
•
Data downloading through the
configuration software or through the
freely distributed download DLL
•
4 input channel types: analog, digital,
remoted and virtual
data loggers
8-Channel Logger
Models DL-1080 & DL-1081
Versatile, powerful and cost
effective data loggers handling
analog, digital and other
types of variables with high
resolution and speed
Special
product
features
▼
R
S
4
85
C O MPAT
IB
L
E
D
C INPUTS
D
IGITAL I/O
UN
IVERSAL
ET
HERNET
P
R
O
GR
AMMA
BL
E
SpEciFicATionS
MODELs
DL-1080 & DL-1081
eLectricaL
Internal Memory
2MB
Power
100 to 240V
ac
, 50/60Hz. 20VA (Max)
Analog Channels Input
Impedance
Thermocouples / Pt100 / Pt1000 / mV: > 2M
Ω
mA: 15
Ω + 1.5V
V: 1.1M
Ω
Accuracy
Thermocouples J, K, T, E and N: 0.2% of the span ± 1°C
Thermocouples R, S and B: 0.2% of the span ± 3°C
Pt100, Pt1000, 0 to 20mA, 4 to 20mA, 0 to 20mV,
0 to 50mV, 0 to 60mV, -20 to 20mV, 0 to 5V and
0 to 10V: 0.15% of the span
Excitation Current
Pt100s: 360μA;
Pt1000s: 320μA
Maximum Pt100/Pt1000
Compensated Cable
Resistance
40
Ω
Digital Inputs
Logic level “0”: from 0 to 0.8V
dc
Logic level “1”: from 3 to 30V
dc
Maximum Input Voltage
30V
dc
Input Current @ 30V
dc
(typical)
3mA
Digital Outputs
Maximum output voltage: 30V
dc
Maximum output current: 200mA
Maximum relay current: 3A @ 250V
ac
; 3A @ 30V
dc
storage Rate
from 1ms to 24 hours
Maximum Channel Logged
100
supported Modbus
Commands
Read Coil Status (01h)
Read Holding Registers (03h)
Write Single Coil (05h)
Write Single Register (06h)
Write Multiple Registers (0Fh)
environMentaL
Operating Temperature
32° to 122°F (0 to 50°C)
Relative Humidity
80% up to 85°F (30°C)
For temperatures higher than 85°F (30°C),
decrease 3% per °C
Altitude
< 2000m
Protection
IP20
Inputs rated CAT II
NEW!
IP
Rated
20