beautypg.com

Isotron, Signal conditioner, Model 2792b – Atec Endevco-2792B User Manual

Page 2: Endevco, Specifications

background image

ENDEVCO

MODEL
2792B

ISOTRON

®

Signal Conditioner

SPECIFICATIONS

INPUTS
TYPE

Single-ended, constant current two-wire system

INPUT IMPEDANCE

> 20 k

EXCITATION CURRENT

4 mA or 10 mA ±10%, set by internal jumper, per channel

COMPLIANCE VOLTAGE

> 18 V. This voltage is the sum of AC and DC components

OUTPUTS
TYPE

Single-ended one side connected to circuit ground

OUTPUT IMPEDANCE

< 10

, in series with at least 40 µF

LINEAR OUTPUT VOLTAGE

10 V pk-pk (3.535 V rms) or greater

LINEAR OUTPUT CURRENT

2.0 mA pk-pk or greater

TRANSFER CHARACTERISTICS
GAIN

0.001 to 99.9 continuously
Adjustable using the full scale Digi switch and THD x1/x10 switch

Accuracy

±1.5% (Including variation with temperature and time)

FREQUENCY RESPONSE

Flat within ±5% from 1 Hz to 30 kHz, reference at 100 Hz

Lower Cutoff Frequency

-5% at 1 Hz, -3dB at 0.3 Hz (typical)

Upper Cutoff Frequency

-5% at 30 kHz, -3dB at 80 kHz (typical)

AMPLITUDE LINEARITY

1% of reading from best-fit straight line approximation

RESIDUAL NOISE

0.4 mV rms maximum for gains 0.5 to 2.0 gain. 2 mV rms maximum for 5.0 to 20.0
or 2.0 or 2 mV rms gains of 5.0 to 20.0 within a 30 KHz bandwidth.

TOTAL HARMONIC DISTORTION

Less than 1% for signals of full scale or less at any gain from .5 to 20.0

CROSSTALK

20.0 mV rms maximum RTO or 1.0 mV rms RTI whichever is greater
Crosstalk specification is valid for the following conditions:
One channel set at 100% of nominal sensitivity and x1 gain
Adjacent channel set at any gain up to 20.0 (eg. 50% nominal sensitivity
and x 10 gain) with a 250

shunt resistor at the input

ENVIRONMENTAL CHARACTERISTICS
TEMPERATURE

Operating +32°F to +122°F (0°C to +50°C)
Storage

-65°F to +185°F (-54°C to +85°C)

HUMIDITY

95% R.H.

POWER
LINE VOLTAGE

100/120/220/240 VAC 50 to 60 Hz (switch selectable)

PHYSICAL
DIMENSIONS

19" rack mounting, 1.73" h x 19.00" w x 9.45" d (44 mm x 483 mm x 240 mm)

WEIGHT

3.97 lbs. (1.8 kg)

FRONT PANEL FEATURES

Power Indicator

LED indicates when power is applied

Gain Switch

Two-position, recessed switch for gain of 1 or 10

Sensitivity Normalization

Four-digit pushbutton switch. See example below

Status Indicators

LED will light green, when the transducer is connected properly. The LED
will light red when the input is shorted. The LED will not light when the input
is open

REAR PANEL FEATURES

Input Connectors

BNC

Output Connectors

BNC

ACCESSORIES
Instruction Manual
EW599

AC POWER CORD

SENSITIVITY NORMALIZATION/ADJUSTMENT:

A four digit push button switch is used to normalize each channel to allow for
variation in nominal sensitivity of the transducer. The switch is set to read the
actual percentage of nominal (desired) sensitivity. For example, if the nominal
(desired) sensitivity is 10.0 mV/g and the actual sensitivity is 10.21 mV/g, set
Percentage of Nominal Sensitivity Switch to 102.1; Gain Switch to x1. If nominal
(desired) sensitivity is 100 mV/g and the actual sensitivity is 10.21 mV/g, set
Percentage of Nominal Sensitivity Switch to 10.21 and Gain Switch to x 10.

Continued product improvement necessitates that Endevco reserve the right to modify these specifications without notice. Endevco maintains a program of con-
stant surveillance over all products to ensure a high level of reliability. This program includes attention to reliability factors during product design, the support of
stringent Quality Control requirements, and compulsory corrective action procedures. These measures, together with conservative specifications have made the
name Endevco synonymous with reliability.

ENDEVCO CORPORATION, 30700 RANCHO VIEJO ROAD, SAN JUAN CAPISTRANO, CA 92675 USA (800) 982-6732 (949) 493-8181 fax (949) 661-7231
www.endevco.com

Email:[email protected]

1100