beautypg.com

Model dt72a, Specifications, Standard equipment – Atec ESI-DT72A User Manual

Page 2: Calibration & technical services

background image

FIRST CORE

SECOND CORE

THIRD CORE

FOURTH CORE

OUTPUT

5

6

GRD

INPUT

GRD

1

2

3

4

11 STEPS

11 STEPS

11 STEPS

11 STEPS

11 STEPS

11 STEPS

10 STEPS

1.1

1.0

0

-0.1

SPECIFIED LINEARITY

AT 1kHz

TYPICAL LINEARITY DEVIATION

AT 1kHz

LINEARITY DEVIA

TION

IN PROPOR

TIONAL P

A

R

T

S

LINEARITY DEVIATION

S E T T I N G

L I N E A R I T Y V E R S U S S E T T I N G

OUTPUT VOLTAGE

INPUT VOLTAGE

– SETTING

10 -7

10 -6

10 -5

10 -4

10 -3

10 -2

10 -1

1

10 -10

10 -9

10 -8

10 -7

10 -6

INPUT IMPED

ANCE MAGNITUDE

IN OHMS

FREQUENCY IN KILOHERTZ

PHASE SHIFT VERSUS FREQUENCY

MAXIMUM PHASE SHIFT

F R E Q U E N C Y I N K I L O H E R T Z

INPUT IMPED

ANCE PHASE ANGLE

IN DEGREES

1M

100k

10k

1k

0.01

0.1

1

10

100

400Hz

500H

100H

1500pF

0.05

0.1

1

10

1

0.1

0.01

0.001

MILLIRADIANS

SECONDS

100

10

1

0.2

1'

400Hz

0.01

0.1

1

10

100

90

0

+90

Model DT72A

D E C A D E

R A T I O

T R A N S F O R M E R

S T A N D A R D

Characteristic

DT72A Specifications

Frequency Range

50 to 20,000 Hz

Maximum rms Input Voltage

0.35f (f in Hz) or 350 volts, whichever is less.

Ratio Range

Maximum

1.111111

Minimum

-0.0111111

Maximum Phase Shift

Approx. 0.05 mrad/kHz for settings above 0.1

Resolution

0.00001% (0.1 ppm)

Maximum Input Current

For best performance, no DC current should be permitted. DC input of 20 µA will decrease AC input voltagerating about 10%
and increase distortion slightly; 200 µA will cause near saturation of core and serious errors.

Terminal Linearity

50 Hz to 1 kHz: ±0.9 ppm (referenced to input) for settings 0.1 to 1.0
[0.9 (10 X setting)

1/2

+ 0.01] ppm for settings 0 to 0.1

Above 1 kHz: multiply by f

2

in kHz

Below 50 Hz: Multiply by 50/f(Hz)

Maximum Effective Series

R: approx. 5 ohms maximum

Output Impedance

L: 30 µH

Input Impedance at 20 V and 400 Hz

Approximately 500 K ohms minimum

Input Inductance

Approximately 100 to 400 H depending on excitation.

Output Current

1 A maximum

Dimensions

19 in (48.3 cm) wide
5-1/4 in (13.3 cm) high
7.1 in (18 cm) deep

Weight

18 lbs. (8.1 kg) net

Standard Equipment

The model DT72A comes with a 7203 instruction manual.

Calibration & Technical Services

For warranty and remedial repair, calibration

services and spare parts, or for additional

information on TEGAM sales and service

offices around the world, contact us at

440-466-6100 (ph) or 440-466-6110 (fx).

Specifications

TEN TEGAM WAY • GENEVA, OHIO 44041

440-466-6100 • FAX 440-466-6110

www.tegam.com • e-mail: [email protected]

YOUR GLOBAL SOURCE FOR TEST

AND MEASUREMENT SOLUTIONS

®

This data sheet was current when it was produced. However, products are constantly being updated and improved. Because of this some differences may occur between
the descriptions herein and the current product. Prices and specifications may be changed without notice.