Demodulated Output Distortion:

<0.3% THD for

50% depth, <0.6% THD for

95% depth.

FM Rejection (at 400 Hz and 1 kHz rates, 50 Hz

to 3 kHz BW):

2

250 kHz to 10 MHz: <0.2% AM for <5 kHz

peak

deviation.
10 MHz to 1300 MHz: <0.2% AM for <50 kHz

peak

deviation.

Residual AM (50 Hz to 3 kHz BW):

<0.01%

rms

.

Supplemental Characteristics

Maximum Depth Resolution:

0.01% for depths

39.99%.

0.1% for depths

40%.

Detectors: peak [+ peak], trough [

peak],

average (rms sine wave calibrated).

Demodulated Output Across an Open Circuit

(600

output impedance):

4

0.1 V/percent when resolution is 0.01%.
0.01 V/percent when resolution is 0.1%.

Frequency Counter

Range:

150 kHz to 1300 MHz.

Sensitivity

:

150 kHz to 650 MHz: 12 mV

rms

(

25 dBm).

650 MHz to 1300 MHz: 22 mV

rms

(

20 dBm).

Accuracy:

Reference accuracy ±3 counts of least

significant digit.

Internal Reference:

Frequency: 10 MHz.

Aging Rate: <1x10

6

/month.

(Optional: <1x10

9

/day)

6

.

Supplemental Characteristics

Modes: Frequency, and frequency error (displays

the difference between the frequency entered
via the keyboard and the actual RF input
frequency).

Sensitivity in Manual Tuning Mode: Approximate

frequency must be entered from keyboard.
0.22 mV

rms

(

60 dBm).

Maximum Resolution:

10 Hz for frequencies <1 GHz.
100 Hz for frequencies

1 GHz.

Internal Reference Accuracy: Overall accuracy

is a function of time base calibration

± aging

rate

± temperature effects ± line voltage effects

± short term stability.

2 Peak residuals must be accounted for in peak readings.

4 For optimum flatness, cables should be terminated with their

characteristic impedance.

5 For peak measurements only, AM accuracy may be affected by

distortion generated by the modulation analyzer. In the worst
case this can decrease accuracy by 0.1% of reading for each 0.1%
of distortion

6 After 30-day warm-up

Standard

Option 002

Aging Rate

<1 x 10

-6

/mo. <1 x 10

-9

/day

Temperature Effects

<2 x 10

-7

/° C <2 x 10

-10

/° C

Line Voltage Effects

<1 x 10

-6

<6 x 10

-10

(+5%,

10% line

voltage change)

Short term stability

<1 x 10

-9

for 1s average

Accuracy:

2

±3% of reading ±1 digit.

Demodulated Output Distortion:

<0.1% THD.

AM Rejection (for 50% AM at 1 kHz rates):

2

<0.03 radians, peak deviation.

Supplemental Characteristics

Modulation Rates: Usable from 20 Hz to 100 kHz

with degraded performance.

Detectors: + peak,

peak, average (rms sine wave

calibrated).

Demodulated Output across an Open Circuit

(600

output impedance):

4

1 V/rad when resolution is 0.001 radian.
0.1 V/rad when resolution is 0.01 radian.
0.01 V/rad when resolution is 0.1 radian.

Amplitude Modulation

Rates:

150 kHz to 10 MHz: 20 Hz to 10 kHz.
10 MHz to 1300 MHz: 20 Hz to 100 kHz.

Depth:

to 99%.

Accuracy:

2, 5

150 kHz to 10 MHz: ±2% of reading ±1 digit,

50 Hz to 10 kHz rates, >5% depth.
±3% of reading ±1 digit, 20 Hz to 10 kHz rates.

10 MHz to 1300 MHz: ±1% of reading ±1 digit,

50 Hz to 50 kHz rates, >5% depth.
±3% of reading ±1 digit, 20 Hz to 100 kHz rates.

Flatness (variation in indicated AM depth for

constant depth on input signal):
10 MHz to 1300 MHz: ±0.3% of reading ±1 digit,
90 Hz to 10 kHz rates, 20 to 80% depth.

Deviation and Maximum Resolution:

0.1 Radian
Resolution

0.01 Radian
Resolution

0.001 Radian
Resolution

400

40

4

200 Hz

1 kHz

10 kHz

20 kHz

Modulation rate

Peak phase deviation (radians)