Technical specifications, Remote programming, General specifications – Atec Agilent-8960 User Manual
Page 3
3
Technical Specifications
These specifications apply to any E5515E or E5515C with serial number
US40410101, GB4041010, MY46010000 or higher.
Specifications describe the test set’s warranted performance and are valid for
the unit’s operation within the stated environmental range unless otherwise
noted. All specifications are valid after a 30-minute warm-up period of
continuous operation.
Supplemental characteristics are intended to provide typical, but non-warranted,
performance parameters that may be useful in applying the instrument. These
characteristics are shown in italics and labeled as “typical”, or “supplemental.”
All units shipped from the factory meet these typical numbers at 25 °C ambient
temperature without including measurement uncertainty.
These are general specifications that apply to an E5515C or E5515E mainframe
over the 25 °C ±5 °C ambient temperature range. Specific test and lab application
specifications can be found in the library on the 8960 Web site:
www.agilent.com/find/8960
Remote programming
GPIB
IEEE Standard 488.2
GPIB help
Pressing the front panel Help key and then any other key will cause the test set to
display the GPIB syntax for that command at the bottom of the front panel display;
pressing the Help key again exits this mode of operation
Remote front panel lockout
Allows remote user to disable the front panel display to improve GPIB measurement
speed
Implemented functions
T6, TE0, L4, LE0, SH1, AH1, RL1, SR1, PP0, DC1, DT0, C0, and E2
General specifications
Dimensions (H x W x D)
235 mm x 425 mm x 629 mm
Weight
31.8 kg
Display
26.7 cm, active matrix, color, liquid crystal
LAN (local area network) port
LAN port (for firmware upgrades only)
LAN 2 port (for high data throughput)
LAN 3 port (for future use)
RJ-45 connector, 100 Mbps
RJ-45 connector, 100 Mbps (for E5515C) or 1000 Mbps (for E5515E)
RJ-45 connector, 1000 Mbps
Operating temperature
0 to +55 °C
Storage temperature
−20 to +70 °C
Power
88 to 135 Vac, 193 to 269 Vac, 50 to 60 Hz, typically 550 VA maximum
Calibration interval
Two years
EMI
Conducted and radiated interference meets CISPR-11
Electrical safety
Complies with CAN/CSA 22.2 No. 61010-1-04, UL Std. 61010-1 (2nd Edition), and IEC
61010-1 (2nd Edition)
Radiated leakage due to RF generator
Typically < 1 μV induced in a resonant dipole antenna one inch from any surface except
the underside and rear panel at set RF generator output frequency and output level of
−40 dBm
Spurious leakage
Typically < 5 μV induced in a resonant dipole antenna one inch from any surface on
the front half of all sides of the instrument at frequencies other than the RF generator
output frequency and output level of −40 dBm with no cable connected to the rear-
panel LAN port
Power consumption
Typically 400 to 450 W continuous