Frequency Modes Gen: 1 Hz to 19,999.9 Hz

Tracking Generator

Swept frequency output is exactly equal to input frequency of spectrum
monitor

Frequency Range: 10 MHz to 999.9999 MHz

Sweep Widths: Fixed - 10 kHz, 100 kHz, 1 MHz per div., and 10-1000 MHz
range (for cable fault)

Output Level: 0.1 µV to 300 mV, continuous

Spectrum Monitor

Frequency Range: 10 to 1000 MHz

Dynamic Range: +0 to -115 dB

RF Attenuator: ≥40 dB in 20 dB steps

Display Range: 70 dB (10 dB/div.) and 8 dB (1 dB/div)

Scan Widths: Fixed - 10, 100 kHz/div, 1, 10 MHz/div., and 0-1000 MHz

Oscilloscope

Frequency Range: DC to 500 kHz

Horizontal Sweep Rate (Calibrated): 10 msec/div, 1 msec/div, 100 µsec/div, 10
µsec/div, ±5%

Signal Generator

Frequency Range: 0.45 to 999.9999 MHz

Level - Range: 0.1 µV to 300 mV rms, continuous

Modulation: FM/AM/CW

General

Approx. Weight: 44 lbs

Dimensions: 8 1/4" H x 15" W x 18" D

Power Sources: 115/230 VAC ± 10%, 50-400 Hz, 90 Watts