Technical specifications, Remote programming, General specifications – Atec Agilent-E5515C User Manual
Page 3
3
Technical Specifications
These specifications apply to an E5515C mainframe for serial number US40410101,
GB4041010, MY46010000 or higher.
Specifications describe the test set’s warranted performance and are valid for the unit’s
operation within the stated environmental range unless otherwise noted. All
specifications are valid after a 30-minute warm-up period of continuous operation.
Supplemental characteristics are intended to provide typical, but non-warranted,
performance parameters that may be useful in applying the instrument. These
characteristics are shown in italics and labeled as “typical”, or “supplemental.” All
units shipped from the factory meet these typical numbers at 25 °C ambient
temperature without including measurement uncertainty.
These are general specifications that apply to an E5515C mainframe over the 25 °C
±5 °C ambient temperature range. Specific test and lab application specifications can
be found in the library on the 8960 website:
www.agilent.com/find/8960
Remote programming
GPIB
IEEE Standard 488.2
GPIB help
pressing the front panel Help key and then any other
key will cause the test set to display the GPIB syntax
for that command at the bottom of the front panel
display; pressing the Help key again exits this mode
of operation
Remote front panel lockout
allows remote user to disable the front panel display
to improve GPIB measurement speed
Implemented functions
T6, TE0, L4, LE0, SH1, AH1, RL1, SR1, PP0, DC1, DT0,
C0, and E2
General specifications
Dimensions (H x W x D)
8.75 x 16.75 x 24.63 inches (222 x 426 x 625 mm), 7
rack spaces high
Weight
66 lbs (30 kg)
Display
10.5 inches (26.7 cm), active matrix, color, liquid crystal
LAN (local area network) port
(for firmware upgrades and
protocol logging only)
RJ-45 connector, 10 base T Ethernet with TCP/IP
support
Operating temperature
0 to +55 °C
Storage temperature
−20 to +70 °C
Power
100 to 240 Vac, 50 to 60 Hz, 550 VA maximum
Calibration interval
two years
EMI
conducted and radiated interference meets CISPR-11
Radiated leakage due to RF
generator
typically < 1 μV induced in a resonant dipole antenna
one inch from any surface except the underside and
rear panel at set RF generator output frequency and
output level of −40 dBm
Spurious leakage
typically < 5 μV induced in a resonant dipole antenna
one inch from any surface on the front half of all sides
of the instrument at frequencies other than the RF
generator output frequency and output level of −40
dBm with no cable connected to the rear-panel LAN
port
Power consumption
typically 400 to 450 W continuous