Operation
photoLab
®
6600 UV-VIS
72
ba75848e01
08/2009
2
Insert the cell with the reference
sample.
The first single measurement for
the Reference absorbance is
carried out.
The following data is displayed as
the result:
– The measured absorbance
from the (last) single
measurement.
– The median from all single
measurements carried out up to
now.
3
If necessary, carry out further
single measurements for the
formation of the median with [Next
meas.]
or
discard the last single
measurement with [Discard].
4
To accept the median value, press
[Apply].
The photometer is ready to
measure.
The reference absorbance is
displayed in the top right corner
during absorbance or
transmission measurement.
Reference absorbance
04/16/07 9:52
To start measurement,
insert cell or press
489 nm
10 mm
Reference absorbance
04/16/07 9:52
Last measured absorbance
0.232
Median
0.232 (1 Measurement(s))
489 nm
10 mm
Next meas.
Discard
Apply
Absorbance
04/16/07 9:52
Reference::
To start measurement,
insert cell or press
489 nm
10 mm
Setup
Wavelength
Transmission
Reference
- Lab 850 BENCHTOP pH METER (73 pages)
- Lab 850, up to serial number 09470314 (61 pages)
- Lab 860 BENCHTOP pH METER (74 pages)
- Lab 870 BENCHTOP pH METER (71 pages)
- Lab 960 (77 pages)
- Lab 970 (71 pages)
- CG 839 (8 pages)
- CG 842 (44 pages)
- CG 843P (72 pages)
- CG 843 (68 pages)
- CG 853P (76 pages)
- CG 853 (74 pages)
- CG 855 (11 pages)
- HandyLab 100 QuickStart (50 pages)
- HandyLab 100 (66 pages)
- HandyLab 200 QuickStart (50 pages)
- HandyLab 200 (52 pages)
- HandyLab 600 QuickStart (50 pages)
- HandyLab 600 (72 pages)
- HandyLab 680 QuickStart (50 pages)
- HandyLab 680 (120 pages)
- HandyLab 700(EX) (128 pages)
- HandyLab 700 (76 pages)
- HandyLab 780 (164 pages)
- handylab LF 11 (44 pages)
- handylab LF 12 (64 pages)
- handylab OX 12 (60 pages)
- handylab multi 12 (82 pages)
- handylab pH 11 (48 pages)
- handylab pH 12 (68 pages)
- handylab pH LF 12 (80 pages)
- IQ SensorNet FDO 70x IQ (SW) (48 pages)
- IQ SensorNet ViSolid 700 IQ (SW) (58 pages)
- MultiLab 4010-2 (112 pages)
- MultiLab 4010-3 (112 pages)
- EBI 40 TK-12 (18 pages)
- TFX 410 (38 pages)
- EBI 20 Serie (20 pages)
- EBI20 TEMPERATURE/HUMIDITY DATALOGGERS (40 pages)
- EBI 20-TF (40 pages)
- TFX 422 (36 pages)
- ST 1000 (42 pages)
- TLC 1598 (2 pages)
- TTX 100/110/120 (28 pages)
- EBI 25 (18 pages)