7 measuring diluted samples, Measuring diluted samples, Section 4.5.7 – Xylem photoLab 6600 UV-VIS User Manual
Page 50

Operation
photoLab
®
6600 UV-VIS
50
ba75848e01
08/2009
4.5.7
Measuring diluted samples
If the concentration of a sample exceeds the measuring range of a method,
you can specifically dilute the sample so that the concentration of the diluted
sample is in the measuring range of the method. Thus a valid measurement
is possible.
After entering the factor for the dilution the meter converts the concentration
to that of the undiluted sample.
Note
Optimum measurement results are achieved if the concentration of the
diluted sample is in the middle of the measuring range of the method after
diluting.
Setting the dilution
Inserting a cell with barcode starts
a measurement.
If a cell without barcode is used:
Select the method manually
(see section 4.5.5).
The photometer is ready to
measure.
Concentration
04/16/07 9:52
Please select method for measuring
or insert a barcoded cell
or insert AutoSelector.
Setup
Method list
Last method
New Method
Concentration
04/16/07 9:52
To start measurement,
insert cell or press
3: A6/25
NH
4
-N
16 mm
0.20 - 8.00 mg/l
Setup
Method list
Citation form
Unit