7 technical data, 1 measurement characteristics, Technical data – Xylem photoLab 6600 UV-VIS User Manual

Page 157: Measurement characteristics, Chapter 7 t, Echnical, Data, En in chapter 7 t, Nical, 7technical data

background image

photoLab

®

6600 UV-VIS

Technical data

157

ba75848e01

08/2009

7

Technical data

7.1

Measurement characteristics

Measuring principle

Single-beam spectrophotometer

Light source

Lamp type

Xenon flashlamp

Durability

5 x 10

8

flashes, corresponding to at least

13000 h in permanent operation

Monochromator

Type

Grating monochromator with step motor

Wavelength range

190 - 1100 nm

Max. scan speed

approx. 3300 nm/min

Wavelengths calibration

Automatic

Accuracy

± 1 nm

Reproducibility

± 0.5 nm
(checkable, e.g. with holmium oxide filter)

Resolution

1 nm

Spectral band width

4 nm

Photometric

measurement

Light sensor

Photo diode

Measuring range

A = -3.300 to A = +3.300

Linearity

< 1 % for A

≤ 2.000

in the range 340 to 900 nm

Accuracy *

– 0.003 A for A < 0.600

– 0.5 % of the reading for 0.600

≤A≤ 2.000

Reproducibility *

±

0.002 at A = 1.000

Resolution

Δ

A = 0.001

Scattered light

< 0.1 % transmission at 340 and 408 nm

* in the range 200nm ... 1000nm