Xylem PH3210 FIELD PH METER User Manual
Page 66

Lists
pH 3210
66
ba75793e02
12/2008
Molality
Molality is the quantity (in Mol) of a dissolved substance in 1000 g
solvent.
ORP voltage
The ORP is caused by oxidizing or reducing substances dissolved in
water, if these substances become effective at an electrode surface
(e. g. a gold or platinum surface).
pH value
The pH is a measure of the acidic or basic effect of an aqueous
solution. It corresponds to the negative decadic logarithm of the molal
hydrogen ions activity divided by the unit of the molality. The practical
pH value is the value of a pH measurement.
Potentiometry
Name of a measuring technique. The signal (depending on the
measured parameter) of the electrode is the electrical potential. The
electrical current remains constant.
Reset
Restoring the original condition of all settings of a measuring system.
Resolution
Smallest difference between two measured values that can be
displayed by a meter.
Slope
The slope of a linear calibration function.
Stability control
(AutoRead )
Function to control the measured value stability.
Standard solution
The standard solution is a solution where the measured value is
known by definition. It is used to calibrate a measuring system.
Test sample
Designation of the test sample ready to be measured. Normally, a test
sample is made by processing the original sample. The test sample
and original sample are identical if the test sample was not processed.
Zero point
The zero point of a pH combination electrode is the pH value at which
the electromotive force of the pH combination electrode at a specified
temperature is zero. Normally, this is at 25 °C.