Example 2 on navigation: setting the date and time – Xylem PH3210 FIELD PH METER User Manual

Page 22

background image

Operation

pH 3210

22

ba75793e02

12/2008

4.2.4

Example 2 on navigation: Setting the date and time

The meter has a clock with a date function. The date and time are
indicated in the status line of the measured value display.
When storing measured values and calibrating, the current date and
time are automatically stored as well.

The correct setting of the date and time and date format is important for
the following functions and displays:

z Current date and time

z Calibration date

z Identification of stored measured values.

Therefore, check the time at regular intervals.

Note
After a fall of the supply voltage (empty batteries), the date and time are
reset to 01.01.2008 00, 00:00 hours.

Setting the date, time

and date format

The date format can be switched from the display of day, month, year
(dd.mm.yyyy) to the display of month, day, year (mm/dd/yyyy or
mm.dd.yyyy).

1

In the measured value display:
Using

_

>/[Menu], open the Storage & config menu.

The instrument is in the setting operating mode.

2

Select and confirm the System / Clock menu with <

▲><▼> and

.
The setting menu for the date and time opens up.

3

Select and confirm the Time menu with <

▲><▼> and

.
The hours are highlighted.

Date format

:

dd.mm.yyyy

Date

:

30.10.2008

Time

:

14:53:40

Clock

22.09.2008

08:00

Back