Ph 2 – Xylem PH3210 FIELD PH METER User Manual

Page 45

background image

pH 3210

Operation

45

ba75793e02

12/2008

Note
For single-point calibration, the instrument uses the Nernst slope
(-59.16 mV/pH at 25 °C) and determines the zero point of the
electrode.

7

Wait for the measurement with stability control to be
completed.
Set the nominal buffer value for the measured temperature with
<

><>.

8

Take over the calibration value for the asymmetry with
.
The calibration display for the next buffer appears.

9

If necessary, terminate the calibration as a single-point
calibration with .
The calibration record is displayed.

pH

7.000

24.8 °C

ConCal

1

22.09.2008

08:00

Asymmetry

pH

2

7.000

24.8 °C

ConCal

22.09.2008

08:00