4 settings for ph and orp measurements, Settings for ph and orp measurements – Xylem PH3210 FIELD PH METER User Manual

Page 32

background image

Operation

pH 3210

32

ba75793e02

12/2008

4.4.4

Settings for pH and ORP measurements

Overview

The following settings are possible for pH and ORP measurements:

z Resolution

z Calibration interval

z Buffers for calibration

z Unit of the temperature

z Automatic stability control

z Unit for slope

z Calibration record (display)

Settings

The settings are made in the measuring menu of the pH/ORP
measurement. To open the settings, activate the relevant measuring
window in the measured value display and press the key
shortly. After completing the settings, switch to the measured value
display with .

Menu item

Possible
setting

Description

Calibration /
Calibration record

-

Displays the calibration
record of the last calibration.

Calibration /Buffer

TEC
NIST/DIN
ConCal
...

Buffer sets to be used for pH
calibration.
More buffers and details,
see section 4.4.5.

Calibration /One
point calibration

Yes
No

Rapid calibration with 1
buffer

Calibration /
Calibration interval

1 ... 999 d

Calibration interval for the
pH electrode (in days).
The meter reminds you to
calibrate regularly by the
flashing sensor symbol in
the measuring window.

Calibration /Unit for
slope

mV/pH
%

Unit of the slope.
The % display refers to the
Nernst slope of
-59.16 mV/pH (100 x
determined slope/Nernst
slope).

Man. temperature

-25 ... +130 °C

Entry of the manually
determined temperature.
For measurements without
temperature sensor only.