4 settings for ph and orp measurements, Settings for ph and orp measurements – Xylem PH3210 FIELD PH METER User Manual
Page 32
Operation
pH 3210
32
ba75793e02
12/2008
4.4.4
Settings for pH and ORP measurements
Overview
The following settings are possible for pH and ORP measurements:
z Resolution
z Calibration interval
z Buffers for calibration
z Unit of the temperature
z Automatic stability control
z Unit for slope
z Calibration record (display)
Settings
The settings are made in the measuring menu of the pH/ORP
measurement. To open the settings, activate the relevant measuring
window in the measured value display and press the
shortly. After completing the settings, switch to the measured value
display with
Menu item
Possible
setting
Description
Calibration /
Calibration record
-
Displays the calibration
record of the last calibration.
Calibration /Buffer
TEC
NIST/DIN
ConCal
...
Buffer sets to be used for pH
calibration.
More buffers and details,
see section 4.4.5.
Calibration /One
point calibration
Yes
No
Rapid calibration with 1
buffer
Calibration /
Calibration interval
1 ... 999 d
Calibration interval for the
pH electrode (in days).
The meter reminds you to
calibrate regularly by the
flashing sensor symbol in
the measuring window.
Calibration /Unit for
slope
mV/pH
%
Unit of the slope.
The % display refers to the
Nernst slope of
-59.16 mV/pH (100 x
determined slope/Nernst
slope).
Man. temperature
-25 ... +130 °C
Entry of the manually
determined temperature.
For measurements without
temperature sensor only.